Parameter

Modes

Range

Description

 

CPAP AutoSet VAuto S ST T PAC iVAPS ASV ASVAuto

 

 

Easy-

ü

On / Off

Enables Easy-Breathe comfort

Breathe

 

Default: On

feature.

(US Only)

 

 

 

 

 

Max

ü

Min–20 cm H2O,

Sets the upper limit of

Pressure

 

0.2 cm H2O

treatment pressure.

 

 

increments

 

 

 

Default:

 

 

 

20 cm H2O

 

Min

ü

4–Max cm H2O,

Sets the lower limit of treatment

Pressure

 

0.2 cm H2O

pressure.

 

 

increments

 

 

 

Default:

 

 

 

4 cm H2O

 

Min EPAP

ü

4–15 cm H2O

Sets the minimum EPAP

 

 

Default:

delivered by the device.

 

 

 

 

 

4 cm H2O

 

Max EPAP

ü

4–15 cm H2O

Sets the maximum EPAP

 

 

Default:

delivered by the device.

 

 

Dependent on Min EPAP.

 

 

15 cm H2O

 

 

 

Options

 

 

 

Climate

ü ü ü ü ü ü ü ü ü ü

Auto / Manual

Sets the type of Climate

Control

 

Default: Auto

Control.

 

 

 

Therapy LED

ü ü ü ü ü ü ü ü ü ü

On / Off

Enables or disables the blue

 

 

Default: Off

LED.

 

 

 

Date

ü ü ü ü ü ü ü ü ü ü

DD Mmm YYYY

Sets the current date or time.

Time

ü ü ü ü ü ü ü ü ü ü

00:00 (24 hr)

If you set a new date or time

that occurs in the past then an

 

 

 

 

 

 

‘Invalid date/time, data exists

 

 

 

for this period’ message is

 

 

 

displayed.

Configuration

 

 

 

Language

ü ü ü ü ü ü ü ü ü ü

English / Français /

Sets the display language.

 

 

Deutsch /

 

 

 

Español /

 

 

 

Português

 

 

 

Default: English

 

Restore

ü ü ü ü ü ü ü ü ü ü

Yes / No

Resets machine default settings

factory

 

 

(except for language, date and

defaults

 

 

time).

Pressure

ü ü ü ü ü ü ü ü ü ü

cm H2O / hPa

Sets pressure unit.

units

 

Default: cm H2O

 

 

 

 

Temperature

ü ü ü ü ü ü ü ü ü ü

°F / °C

Sets temperature unit.

units

 

Default: °F

 

 

 

 

Height units

ü ü ü ü ü ü ü ü ü ü

cm / inches

Sets height unit.

 

 

Default: inches

 

18