Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device.

Problem/Possible cause

Solution

No display

 

Power is not connected.

Ensure the power cord is connected and the power outlet (if

 

available) is on.

The DC plug is partially inserted into

Fully insert the DC plug.

the back of the device or inserted

 

too slowly.

 

The S9 VPAP Tx and H5i are not

Ensure that the S9 VPAP Tx and H5i are securely attached.

connected correctly.

 

Insufficient air delivered from the device

Air filter is dirty.

Replace air filter.

Air tubing is not connected properly.

Check air tubing.

Air tubing is blocked, pinched or

Unblock or free the air tubing. Check the air tubing for

punctured.

punctures.

Mask and headgear are not

Adjust position of mask and headgear.

positioned correctly.

 

Incorrect air tubing selected.

If you are using the SlimLine, Standard or 3 m air tubing ensure

 

that you have the correct air tubing selected via the menu.

The H5i flip lid is not latched

Close the flip lid ensuring that it clicks into place.

correctly.

 

The H5i flip lid seal is not fitted

Make sure the flip lid seal is facing the right way up and fitted

correctly.

securely.

Mask vents might be blocked.

Check if you have sufficient venting. Unblock mask vents if

 

necessary.

EPAP may be set too low.

Adjust EPAP settings.

There is excessive leak.

Adjust position of mask and headgear.

Pressure rises inappropriately

 

Talking, coughing or breathing in an

Avoid talking with a nasal mask on, and breathe as normally as

unusual manner.

possible.

Mask cushion is buzzing against the

Adjust the headgear.

skin.

 

Cushion seated incorrectly causing

Adjust headgear or re-fit cushion.

excessive leak.

 

Displays message: Heated tube fault, replace tube

Device has been left in a hot

Allow to cool before re-use. Disconnect the power cord and

environment.

then reconnect it to restart the device.

There is a fault in your ClimateLine

Discontinue using your ClimateLine or ClimateLineMAX heated

or ClimateLineMAX heated air tubing.

air tubing and contact your service provider. Use SlimLine,

 

Standard or 3 m air tubing in the interim.

Displays message: High temperature fault, refer to user manual

Device has been left in a hot environment.

Allow to cool before re-use. Disconnect the power cord and then reconnect it to restart the device.

25 English