ResMed S9 VPAP Tx manual Clinical setup menu parameters, Parameter Modes, Range Description, Ü ü

Models: S9 VPAP Tx

1 37
Download 37 pages 51.06 Kb
Page 20
Image 20

Clinical setup menu parameters*

Parameter

Modes

CPAP

AutoSet

VAuto

S

ST

T

PAC

iVAPS

ASV

ASVAuto

Range

Description

Settings

Circuit

Mask type

ü ü ü ü ü ü ü ü ü ü

Full Face / Nasal

Selects the type of mask used

 

 

/ Pillows / Nasal

by the patient.

 

 

Ultra / Pediatric

 

 

 

Default: Full Face

 

Tube type

ü ü ü ü ü ü ü ü ü ü

SlimLine /

Shows the type of air tubing

 

 

Standard / 3 m

used by the patient.

 

 

Default: Standard

Not displayed if ClimateLine or

 

 

 

ClimateLineMAX is connected

AB filter

ü ü ü ü ü ü ü ü ü ü

No / Yes

Enables or disables antibacterial

 

 

Default: No

filter.

 

 

Not displayed if H5i is

 

 

 

 

 

 

connected.

Ext.

ü ü ü ü ü ü ü ü ü ü

No / Yes

Enables or disables an external

humidifier

 

Default: No

humidifier.

 

 

 

Therapy

 

 

 

Mode

ü ü ü ü ü ü ü ü ü ü

CPAP / S / ST /

Sets the therapy mode available

 

 

T / PAC / iVAPS /

on the device.

 

 

ASVAuto / ASV /

 

 

 

AustoSet / VAuto

 

 

 

Default: CPAP

 

Set pressure

ü

4–20 cm H2O

Sets the fixed treatment

 

 

Default:

pressure.

 

 

 

 

 

4 cm H2O

 

EPR Level

ü ü

Off / 1 / 2 / 3

Sets the EPR value.

 

 

Default: 3

 

Height

ü

44-100 inches

Body height needed for the

 

 

(110–250 cm)

dead space determination.

 

 

Default: 70 inches

 

 

 

(175 cm)

 

Target Pt

ü

8–30 BPM

The rate input to the iVAPS

Rate

 

Default: 15 BPM

algorithm. This should be set at

 

 

the patient’s actual respiratory

 

 

 

 

 

 

rate.

Target Va

ü

1–30 L/min

Used to determine the amount

 

 

Default: 5.2 L/min

of pressure support required by

 

 

the iVAPS algorithm.

 

 

 

IPAP

ü ü ü ü

4–30 cm H2O

Sets the pressure which will be

 

 

Default:

delivered to the patient when

 

 

the device is triggered into

 

 

10 cm H2O

 

 

inspiration.

 

 

 

*Default settings may differ from country to country

 

 

16