Managing Climate Control

Designed to be ideal for most patients, Climate Control Auto enables the automatic delivery of a constant value of absolute humidity while protecting against rainout.

To allow for increased flexibility, Climate Control can be turned to Manual in the Setup menu. Setting Climate Control to Manual disables the automatic control of humidity and allows humidity and temperature levels to be set independently. However, rainout protection is not provided when Climate Control is set to Manual.

Mode

Humidity

 

Temperature

Setting range

Default settings

Setting range

Default settings

 

Climate Control – Auto

 

 

Climate Control

 

 

Constant

 

Off*, 60ºF–86ºF

80ºF (27ºC)

 

absolute humidity

 

 

(16ºC–30ºC)

 

 

(depending on

 

 

 

 

 

temperature

 

 

 

 

 

setting)

 

 

 

 

Climate Control –

Humidity level

 

Heated tube

Manual

Off–6.0

 

3

Off, 60ºF–86ºF

80ºF (27ºC)

 

 

 

(0.5 increments)

 

 

(16ºC–30ºC)

 

*When the temperature setting is set to Off the tube will not heat the air, nor will the humidifier heat the water to add humidity to the air.