Sound:

DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:1996

Sound pressure level (CPAP mode)

with SlimLine air tubing: 26 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2007

with Standard air tubing: 27 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2007

with either SlimLine or 28 dBA with uncertainty of 2 dBA as measured Standard air tubing and H5i: according to ISO 17510-1:2007

Sound power level (CPAP mode)

with SlimLine air tubing: 34 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2007

with Standard air tubing: 35 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2007

with either SlimLine or 36 dBA with uncertainty of 2 dBA as measured Standard air tubing and H5i: according to ISO 17510-1:2007

Alarm volume

Low (nominal 56 dBA), Medium (nominal 68 dBA), High (nominal 80 dBA)

settings

 

Physical

Dimensions (L x W x H): 6.0” x 6.8” x 3.4” (153 mm x 172 mm x 86 mm)

 

Weight: 2.30 lbs (1.045 kg)

 

Housing construction: Flame retardant engineering thermoplastic

 

Air outlet: 22 mm conical air outlet (complies with ISO 5356-1:2004)

Air filter

Standard: Polyester non-woven fiber

 

Hypoallergenic: Acrylic and polypropylene fibers in a polypropylene carrier

Supplemental

Recommended maximum supplemental oxygen: 15 L/min (CPAP, S, T, ST, ASV and

oxygen

PAC modes), 4 L/min (AutoSet, VAuto, iVAPS and ASVAuto modes)

Pneumatic flow path

 

1

2

3

3

 

 

4

1.

Flow sensor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2.

Blower

 

 

 

 

 

 

 

 

3.

Pressure sensor

 

 

 

 

 

 

 

 

4.

Mask

 

 

 

 

 

 

 

 

5.

Air tubing

 

 

 

 

 

 

 

 

6.

H5i

 

 

 

 

 

 

 

 

7.

Alarm module

9

8

 

 

7

6

5

 

8.

Device

 

 

 

9.

Inlet filter

28

Page 32
Image 32
ResMed S9 VPAP Tx manual Pneumatic flow path