Sanyo MDF-U730M Temperature Setting, Key Lock Function, Description of operation, Key operated

Models: MDF-U730M

1 32
Download 32 pages 24.59 Kb
Page 14
Image 14
TEMPERATURE SETTING

TEMPERATURE SETTING

Table 1 shows the basic operation method. Perform key operations in the sequence indicated in the table. The example in the table is based on the assumption that the freezer temperature is -20oC.

Note: The unit is set at the factory that the chamber temperature -30oC.

Table 1 Basic operation sequence (Example: Chamber temperature -25oC)

 

Description of operation

Key operated

Indication after operation

 

 

 

The current chamber temperature is

1

Connect the power source and turn

----

displayed. Chamber temperature

on the power switch.

higher than 1oC to be displayed ”HI”

 

 

 

lower than -41oC to be displayed “LO”.

2

Press set key.

SET

The first digit of the temperature

indicator flashes.

 

 

 

 

 

 

Pressing the digit shift key leads the

 

Set the temperature to -25 with the

 

flash of the second digit.

3

digit shift key and the numerical

 

Pressing the numerical value shifts

 

 

value shift key.

 

key down the figure of the current

 

 

 

digit to 2.

 

 

 

The value is stored in memory and

4

Press set key.

SET

the current chamber temperature

 

 

 

is displayed.

Note:

The setting mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation. In this case, any setting before pressing the set key (SET) is not memorized.

The freezer temperature can be set in the range between -18 and -35oC. Remember that the guaranteed temperature with no load at an ambient temperature of 35oC is -30oC

KEY LOCK FUNCTION

This unit is provided with the key lock function. When the key lock is ON, change of temperature setting through the key pad is not available. The key lock is set in OFF at the factory.

Display

Mode

Function

L o 0

Key lock is OFF

Enable to change of temperature setting

L o 1

Key lock is ON

Disable to change of temperature setting

Table 2 Procedure for key lock setting (change from key lock OFF to key lock ON)

 

Description of operation

Key operated

1

 

----

 

 

 

2

Press digit shift key for 5 seconds.

 

 

 

 

3

Press numerical value shift key and

 

scroll the figure to 1.

 

 

 

4

Press set key.

SET

 

 

 

Indication after operation The current chamber temperature is displayed.

The right digit is flashed.

When pressed, the figure of settable digit increases.

The key lock is set to ON.

The current chamber temperature is displayed.

Note:

The key lock only works with the chamber temperature setting and the defrost key (DEF).

13

Page 14
Image 14
Sanyo MDF-U730M Temperature Setting, Key Lock Function, Basic operation sequence Example Chamber temperature -25oC