<MTR-G85>

TEMPERATURE RECORDER (OPTION)

<MTR-G85>

1.Attach the recorder to the mounting kit (MTR-S7) by following the procedure shown in the manual enclosed with the mounting kit.

2.Remove the screw on the cover for the recorder space on the upper front of the freezer. Open the cover. Fig. 7

3.Join the wire in the unit compartment with the recorder wire by each wire connector. See Fig. 8

4.The sensor of the temperature recorder is put in the chamber through the access port in the top of the product.

5.A recorder sensor is to do fix with a supplied nylon

clip the inside wall.

See fig. 5.

6.A mounting kit and a temperature recorder are

installed on the freezer.

See fig. 9

7.Operate the freezer until the chamber temperature gets to the set temperature. Check the recorded temperature and chamber temperature displayed on the control panel. Adjust the zero adjustment volume on the temperature recorder so that the recorded temperature can corresponds with the displayed temperature if they are not compliance each other