Sanyo MDF-U730M instruction manual Alarms & Safety Functions, Alarms and safety functions

Models: MDF-U730M

1 32
Download 32 pages 24.59 Kb
Page 17
Image 17
ALARMS & SAFETY FUNCTIONS

ALARMS & SAFETY FUNCTIONS

This unit has the alarms and safety functions shown in Table 4, and also self diagnostic functions.

Table 4 Alarms and safety functions

Alarm & Safety

Situation

Indication

Buzzer

Safety operation

High temperature

If the chamber temperature

Alarm lamp is flashed.

Intermittent tone after

Remote alarm is

deviates by +10oC (factory set,

Temperature indicator is

activated after a delay

alarm

changeable) from setting.

flashed.

a delay 15 minutes.

of 15 minutes.

 

 

Low temperature

If the chamber temperature

Alarm lamp is flashed.

Intermittent tone after

Remote alarm is

deviates by -10oC (factory set,

Temperature indicator is

activated after a delay

alarm

changeable) from setting.

flashed.

a delay 15 minutes.

of 15 minutes.

 

 

 

In the case of power failure.

 

 

 

Power failure alarm

When power switch is turned OFF.

Alarm lamp is flashed.

Intermittent tone

Remote alarm is

When the power to the unit is

activated.

 

 

 

 

disconnected.

 

 

 

Door alarm

When the door is open.

Door check lamp is lit.

Intermittent tone after

-----

a delay of 2 minutes.

 

 

 

 

 

If a key operation is not performed

Chamber temperature is

 

Setting mode is

Auto return

for about 90 seconds in each setting

-----

displayed.

canceled.

 

mode.

 

 

 

 

 

Key lock

When the key lock is ON (L1).

-----

-----

Key input is unable.

 

 

 

 

 

 

 

Alarm lamp is flashed.

 

Remote alarm is

 

If the thermal sensor goes

 

activated.

 

E1 and LO are displayed

Intermittent tone

 

open circuit.

Operation is keeping

 

alternately.

 

Thermal sensor

 

 

continuously.

 

 

 

Abnormality

 

Alarm lamp is flashed.

 

Remote alarm is

 

If the thermal sensor goes

 

activated.

 

E1 and HI are displayed

Intermittent tone

 

short circuit.

Operation is keeping

 

alternately.

 

 

 

 

continuously.

 

 

 

 

 

 

 

 

Remote alarm is

 

In case of failure of compressor

 

 

activated.

Compressor temp.

cooling fan motor.

-----

Intermittent tone

Compressor stops

Abnormality

In case of abnormal high ambient

running when the

 

 

 

temperature.

 

 

temperature about

 

 

 

 

95oC.

Note:

The alarm can be canceled by pressing the alarm buzzer stop key (BUZZER), but the remote alarm cannot be silenced.

After a power failure, the unit will resume operation with the set value that was in place before power failure occurred.

16