REPLACEMENT OF BATTERY

REPLACEMENT OF BATTERY

Location of a nickel-metal-hydride battery

This unit is provided a nickel-metal-hydride battery for the power failure warning device. The battery is located in the battery mounting plate inside the cover on the upper side. (Fig. 1)

The high voltage components are enclosed in the front cover. The cover should be removed by a qualified engineer or a service personnel only to prevent the electric shock.

Removal of nickel-metal-hydride battery

1.Turn off the power switch, battery switch and disconnect the power supply plug.

2.Lift the front cover. (Fig. 2)

3.Remove 2 screws fixing the battery with a screw driver. (Fig. 3)

4.Disconnect the connector and take out the battery. (Fig. 4)

5.Follow the procedure for recycling or proper disposal.

Electrical box

 

 

 

Battery

 

Fig. 1

 

Fig. 2

Battery

Battery

 

Battery

fixture

 

fixture

 

 

 

Screw

 

Connector

Fig. 3

Battery

Fig. 4

 

 