Sanyo MDF-U730M instruction manual Temperature Recorder Option, Clip Big

Models: MDF-U730M

1 32
Download 32 pages 24.59 Kb
Page 27
Image 27
Clip (Big)

TEMPERATURE RECORDER (OPTION)

5.Remove 2 rubber caps (outside and inside) on the access port on the back of the freezer and also remove the insulation in the port. Fig. 4

6.A recorder sensor is to do fix with supplied nylon

clips the inside wall by screw.

See fig. 5.

7.Replace the insulation into the access port and cover the port completely with the rubber caps. Seal

the access port with a silicon

See fig. 6.

8.Operate the freezer until the chamber temperature gets to the set temperature. Check the recorded temperature and chamber temperature displayed on the control panel. Adjust the zero adjustment volume on the temperature recorder so that the recorded temperature can corresponds with the displayed temperature if they are not compliance each other

Access port

Fig. 4

Clip (Big)

Fig. 5

 

 

Fig. 6

26