Sanyo MDF-U730M instruction manual Remote Alarm Terminal, N.C. Com N.O

Models: MDF-U730M

1 32
Download 32 pages 24.59 Kb
Page 15
Image 15
REMOTE ALARM TERMINAL

REMOTE ALARM TERMINAL

N.C. COM N.O.WARNING

Always disconnect the power supply cord before connecting an alarm device to the remote alarm terminal.

The terminal of the remote alarm is installed at the top of the unit. The alarm is outputted from this terminal between terminal 1 and terminal 2. Contact capacity is DC 30 V, 2 A.

 

between COM. and N.O.

between COM. and N.C.

At normal

Open

Close

At abnormal

Close

Open

Battery switch

1

2

3

N.C. COM N.O.

Note:

The remote alarm is silenced by pressing the alarm buzzer stop key (BUZZER) as the remote alarm is operated in conjunction with the buzzer except for the power failure alarm and alarm test. The contact closed in the of power failure.

14

Page 15
Image 15
Sanyo MDF-U730M instruction manual Remote Alarm Terminal, N.C. Com N.O