12.Dilute the remaining H2O2 reagent in the H2O2 Generator with a large volume of water and dispose of it. Rinse and wash the H2O2 Generator with distilled water. Then keep the H2O2 Generator in a clean environment outside of the chamber.

Note: After H2O2 decontamination, cover the connector on the chamber side with the connector cap.

13.After H2O2 decontamination, surplus H2O2 liquid will remain at the bottom of the chamber and in the bottom part of the H2O2 Generator duct. This solution contains H2O2 at a low concentration (about 1% or less), so put on protective glasses and rubber gloves and wipe it up with a non-woven cloth.

14.Ventilate the chamber sufficiently and place all the attachments back into the chamber.

Note: Be sure to attach the connector cap.

CAUTION

The electric lock will remained locked if power fails during H2O2 decontamination. After the power comes back ON, the H2O2 gas resolution process will start execution and finish automatically. Execute the decontamination again because the decontamination will be incomplete.

WARNING

Never open the door by unlocking it with the unlock key during H2O2 decontamination or H2O2 gas resolution with the UV lamp. H2O2 gas leakage is potentially harmful to health.

Precautions when handling H2O2 decontamination reagent

Observe the following precautions when handling the H2O2 Decontamination Reagent (MCO-H2O2).

zHandling Precautions

ξWear protective equipment, such as protective glasses and rubber gloves.

ξDo not use fire in the area where the reagent is being handled.

ξDo not leave any reagent in the container after it has been used or while it is being used.

ξDo not place inflammable or combustible materials near the area where the reagent is handled.

zPrecautions for Storage

ξAlways close the container cover securely to prevent impurities from becoming mixed in the reagent.

ξCheck the container to make sure that there is no damage, corrosion, or cracking.

ξStore the container with the inlet facing upwards, and make sure that the container will not tip or be knocked over.

ξIf possible, store the reagent in a location where water is available and where the floor can be easily washed.

zPrecautions for Disposal

ξDispose according to the rule in your country.