AUTOMATIC CO2 DENSITY CALIBRATION
To calibrate the CO2 density, an Automatic CO2 Standard Gas Calibration Kit (MCO-SG) is available. By
connecting a CO2 standard gas cylinder that has a density that is the same as the set CO2 density (usually
5% CO2), the CO2 sensor can be calibrated to maintain the accuracy of the CO2 density display as
necessary.
Ɣ Setting connected standard CO2 gas density
1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER
Key.
2. Press the Down Cursor Key from the Select Tools (1/2) Display to display the Select Tools (2/2) Display,
and then select Standard Gas Setting, press the MENU Key to display the menu, select OK, and press
the ENTER Key.