The buzzer can be stopped by pressing the BUZZER Key, but remote alarms will not stop. In addition, a buzzer that sounds because of the upper limit temperature alarm cannot be stopped by pressing the BUZZER Key. Err01 recovery is achieved automatically by correctly connecting the gas and pressing the BUZZER Key to stop the buzzer. In addition, when the MCO-21GC is installed and the gas supply is switched to the reserve gas cylinder, press the BUZZER Key and replace the gas cylinder. If an error from Err05 to Err18 occurs (except for the upper limit temperature alarm in Err13 to Err16), then consult a Sanyo sales representative or agent. Table 2: Alarm and Safety Functions for H2O2 Decontamination

Alarm and safety

function

Condition Display Buzzer Safety operation

The H2O2 Generator is not
connected.
H2O2 Decon System Error is displayed
in the screen center.
Err31:H2O2 Unit No Connect is
displayed in the message display field.
----- Decontamination mode is
cancelled.
There is no H2O2 solution or
the H2O2 level sensor has
failed (or is disconnected).
H2O2 Decon System Error is displayed
in the screen center.
Err32: H2O2 Level is Low is displayed in
the message display field.
----- Decontamination mode is
cancelled.

System check error

at start of H2O2

decontamination

A door is not closed. H2O2 Decon System Error displayed in
the screen center.
Err33: Door is Open is displayed in the
message display field.
----- Decontamination mode is
cancelled.
The volume of H2O2 generated
is abnormal.
Err34:H2O2 Volume is displayed in the
message display field.
Intermittent
tone
Moves to UV analysis.
Remote alarm operates.

Failure during H2O2

decontamination

Power was interrupted during
H2O2 decontamination.
After power is restored, Err35: Power
Failed is displayed in the screen
message display field.
After completion, Decon Failed is
displayed in the screen center.
Intermittent
tone
Electrically locked during power
interruption. After power is
restored, moves to UV analysis.
Remote alarm operates.
The UV lamp failed during
H2O2 gas analysis by UV.
Err36: UV Lamp Failure is displayed in
the message display field.
Intermittent
tone
UV analysis time is extended.
Remote alarm operates.

Failure during H2O2

gas analysis by UV

Power was interrupted during
H2O2 gas analysis by UV.
After power is restored, Err37: Power
Failed is displayed in the message
display field.
----- Electrically locked during power
interruption. After power is
restored, UV analysis is
repeated.
Table 3: Alarm and Safety Functions for Automatic Standard Gas Calibration

Alarm and safety

function

Condition Display Buzzer Safety operation

System check error

at start of automatic

standard gas

calibration

The gas pressure for the
standard gas calibration CO2
gas line is less than the
specified pressure.
Std Gas Calib Error is displayed in the
screen center.
Err41: CO2 Std Gas Empty is displayed
in the message display field.
Intermittent
tone
The standard gas calibration
mode is cancelled.
Remote alarm operates.

Failure during

automatic standard

gas calibration

The gas pressure for the
standard gas calibration CO2
gas line is less than the
specified pressure.
Std Gas Calib Error is displayed in the
screen center.
Err42: CO2 Std Gas Empty is displayed
in the message display field.
Intermittent
tone
Automatic calibration is
cancelled.
Remote alarm operates.
Note For models with electric locking, an unlock key is provided in order to unlock the door during a power interruption or in case the electric lock fails. Always store this key in a safe place. It is recommended that you make a note of the key symbol and number in case the key is lost.
WARN ING
Do not use the unlock key to unlock the door during H2O2 decontamination or during H2O2 gas analysis by UV. Doing so may cause harm to health from H2O2 gas leakage.