Cal=Trak calibrator. See Chapter
Any service to the Sierra Cal=Trak must be performed by Sierra maintenance personnel.
Current service or calibration information and pricing can be received by contacting Sierra Instruments, Inc. at any of the locations listed on the front cover. Alternatively, you can contact us via our web site www.sierrainstruments.com.
11.0Quality Assurance
11.1Leak Test Procedure
A quality assurance
Since the leak test may take several hours, it is recommended to plug the charger into the base during the leak test to prevent the unit from turning off.
To Initiate the Leak Test
1Place the Cal=Trak on a flat,
2While in the “Main Menu” install the leak test cable (included with the Cal=Trak base unit) between the
Run
Main Menu
Setup Leak Test
3
4
5
6
7
From the “Main Menu” remove the cell and select “Leak Test”
After a leak test is initiated, the display will read:
Place the leak test plug over either the inlet or the outlet port fitting on the Cal=Trak flow cell. Invert the Cal=Trak flow cell
Push any key to commence the leak test. At this point the Cal=Trak will time the descent of the piston. The leak test may take up to 6 hours to complete. The display will read:
8A If the test is completed successfully, the display will read:
24