5-6. USING THE CONTINUOUSLY RECORDED DISC

MDS-NT1

5-6. USING THE CONTINUOUSLY RECORDED DISC

*This disc is used in focus bias adjustment and error rate check. The following describes how to create a continuous recording disc.

1.Insert a disc (blank disc) commercially available.

2.Check the [JOG UP] or [JOG DOWN] and display “CREC 2MODE” (C37).

3.Click the [ENTER/YES] again to display “CREC 2MID”. Display “CREC 2 (0300)” and start to recording.

4.Complete recording within 5 minutes.

5.Click the [MENU/NO] and stop recording .

6.Click the [EJECT] and remove the disc.

The above has been how to create a continuous recorded data for the focus bias adjustment and error rate check.

Note: Be careful not to apply vibration during continuous recording.

5-7. CHECKS PRIOR TO REPAIRS

These checks are performed before replacing parts according to “approximate specifications” to determine the faulty locations. For details, refer to “Checks Prior to Parts Replacement and Adjust- ments” (see page 10).

5-7-1. Temperature Compensation Offset Check

When performing adjustments, set the internal temperature and room temperature to 22 to 28ºC.

Checking Procedure:

1.Click the [JOG UP] or [JOG DOWN] to display “TEMP CHECK” (C12).

2.Click the [ENTER/YES].

3.“T=@@(##) [OK” should be displayed. If “T=@@ (##) [NG” is displayed, it means that the results are bad.

(@@ indicates the current value set, and ## indicates the value written in the non-volatile memory)

5-7-2. Laser Power Check

Before checking, check the Iop value of the optical pick-up. (Refer to 5-9. Recording and Displaying the Iop Information (see page 31)

Connection:

laser

power meter

Optical pick-up objective lens 5-7. CHECKS PRIOR TO REPAIRS

digital voltmeter

BD board

 

 

CN105 pin 1 (I +3 V)

+

CN105 pin 2

(Iop)

Checking Procedure:

1.Set the laser power meter on the objective lens of the optical

pick-up.

Connect the digital volt meter to CN105 pin 1 (I+3V) and CN105 pin 2 (Iop).

2.Then, click the [JOG UP] or [JOG DOWN] and display “LDPWR CHECK” (C13).

3.Click the [ENTER/YES] once and display “L 0.90 mW $ 5-7-1. Temperature Compensation Offset CheckChecking Procedure:5-7-2. Laser Power CheckConnection:”. Check that the reading of the laser power meter become speci- fied value.

Specified Value:

Laser power meter reading : 0.84 to 0.92 mW (KMS-262A)

0.90to 0.96 mW (KMS-262E)

4.Click the [ENTER/YES] once more and display “L 8.40 mW $

Specified Value:5-7-3. Iop Compare(KMS-262A) or “L 8.65 mW $ Manual backgroundManual background(KMS-262E). Check that the reading the laser power meter and digital volt meter sat- isfy the specified value.

Specified Value:

Laser power meter reading : 8.4 ± 0.3 mW (KMS-262A)

8.65± 0.25 mW (KMS-262E)

Digital voltmeter reading : Optical pick-up displayed value ± 10%

(Optical pick-up label)

KMS262A

(For details of the method for

20101

checking this value, refer to “6-9.

Recording and Displaying the Iop

B0825

Information”)

 

 

Iop = 82.5 mA in this case

Iop (mA) = Digital voltmeter reading (mV)/1 ()

5.Click the [MENU/NO] and display “LDPWR CHECK” (C13) and stop the laser emission.

(The [MENU/NO] is effective at all times to stop the laser emis- sion)

Note: After step 4, each time the [ENTER/YES] is clicked, the display will be switched between “L 0.70 mW $ Manual backgroundManual background ” and “L 7.50 mW $ Manual backgroundManual background ”, Nothing needs to be performed here.

Checking Location: BD board (see page 35)

5-7-3. Iop Compare

The current Iop value at laser power 8.4 mW output and reference Iop value (set at shipment) written in the nonvolatile memory are compared, and the rate of increase/decrease will be displayed in percentage.

Note: Perform this function with the optical pick-up set at room temperature.

Procedure:

1.Click the [JOG UP] or [JOGDOWN] to display “Iop Compare” (C27).

2.Click the [ENTER/YES] and start measurements.

3.When measurements complete, the display changes to “± xx%yy”.

xx is the percentage of increase/decrease, and OK or NG is displayed at yy to indicate whether the percentage of increase/ decrease is within the allowable range.

4.Click the [MENU/NO] to end this mode.

28