MDS-NT1
5-8. INITIAL SETTING OF ADJUSTMENT VALUE
Note:
Mode which sets the adjustment results recorded in the
If initial setting is performed, perform all adjustments again excluding the temperature compensation offset adjustment.
For details of the initial setting, refer to
Setting Procedure:
1.Click the [JOG UP] or [JOG DOWN] to display “ADJ CLEAR” (C28).
2.Click the [ENTER/YES]. “Complete!” will be displayed mo- mentarily and initial setting will be executed, after which “ADJ CLEAR” (C28) will be displayed.
5-9. RECORDING AND DISPLAYING THE Iop
INFORMATION
The Iop data can be recorded in the
Recording Procedure:
1.Click the [JOG UP] or [JOG DOWN] to display “Iop Write” (C05), and click the [ENTER/YES].
2.The display becomes “Ref=@@@.@” (@ is an arbitrary num- ber) and the numbers which can be changed will blink.
3.Input the Iop value on the optical
To select the number : Click the [JOG UP] or [JOG DOWN].
To select the digit : Click the [PUSH].
4.When the [ENTER/YES] is clicked, the display becomes “Measu=@@@.@” (@ is an arbitrary number).
5.As the adjustment results are recorded for the 4 value. Leave it as it is and click the [ENTER/YES].
6.“Complete!!” will be displayed momentarily. The value will be recorded in the
Display Procedure:
1.Click the [JOG UP] or [JOG DOWN] to display “Iop Read” (C26), and click the [ENTER/YES].
2.“@@.@/##.#” is displayed and the recorded contents are dis- played.
@@.@ indicates the Iop value on the optical
3.To end, click the [PUSH] or [MENU/NO] to display “Iop Read” (C26).
5-10. TEMPERATURE COMPENSATION OFFSET ADJUSTMENT
Save the temperature data at that time in the
Note:
1.Usually, do not perform this adjustment.
2.Perform this adjustment in an ambient temperature of 22 ˚C to 28 ˚C. Perform it immediately after the power is turned on when the internal temperature of the unit is the same as the ambient temperature of 22 ˚C to 28 ˚C.
3.When D101 has been replaced, perform this adjustment after the tem- perature of this part has become the ambient temperature.
Adjusting Procedure:
1.Click the [JOG UP] or [JOG DOWN] to display “TEMP AD- JUST” (C03).
2.Click the [ENTER/YES] to select the “TEMP ADJUST” mode.
3.“TEMP = [OK” and the current temperature data will be displayed.
4.To save the data, click the [ENTER/YES].
When not saving the data, click the [MENU/NO].
5.When the [ENTER/YES] is clicked, “TEMP = SAVE” will be displayed and turned back to “TEMP ADJUST” (C03) dis- play then. When the [MENU/NO] is clicked, “TEMP ADJUST” (C03) will be displayed immediately.
Specified Value:
The “TEMP = ” should be within “E0 - EF”, “F0 - FF”, “00 - 0F”, “10 - 1F” and “20 - 2F”.
5-11. LASER POWER ADJUSTMENT
Check the Iop value of the optical
Connection:
laser
power meter
Optical
digital voltmeter
BD board |
|
|
CN105 pin 1 (I +3 V) | + | |
CN105 pin 2 | (Iop) | – |
Adjusting Procedure:
1.Set the laser power meter on the objective lens of the optical
Connect the digital voltmeter to CN105 pin 1 (I+3V) and CN105 pin 2 (Iop) on the BD (MD) board.
2.Click the [JOG UP] or [JOG DOWN] to display “LDPWR ADJUST” (C04).
(Laser power : For adjustment)
3.Click the [ENTER/YES] once to display “L 0.90 mW $ ”.
4.Click the [JOG UP] or [JOG DOWN] so that the reading of the laser power meter becomes specified value. Click the [ENTER/YES] after setting the range knob of the laser power meter, and save the adjustment results. (“LD SAVE $ ” will be displayed for a moment)
Specified Value:
Laser power meter reading: 0.85 to 0.91 mW
0.90to 0.95 mW
5.Then “L 8.40 mW $ ”
31