SPECIFICATIONS

 

 

DEVILBISS 4-LITER SERIES

 

 

 

 

DEVILBISS 5-LITER SERIES

 

 

 

 

 

 

 

 

 

 

 

Catalog Number

515UK

 

515DS, 515DZ

 

 

515KS, 515KZ, 515NS

 

 

 

 

 

 

 

 

 

 

 

Delivery Rate (Lower delivery rates

1 to 4 LPM

 

 

 

 

1 to 5 LPM

 

 

available for low flow applications)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Maximum Recommended Flow

4 LPM

 

 

 

 

5 LPM

 

 

 

 

 

 

 

 

 

 

 

 

Outlet Pressure

 

 

8.5 psig (58.6 kPa)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Electrical Rating

230 V, 50 Hz, 1.4 Amp

 

115 V, 60 Hz, 4.2 Amp

 

 

220-230 V~, 50 Hz, 1.7 Amp

 

 

 

 

 

 

 

 

 

230 V~, 60 Hz, 1.9 Amp

 

 

 

 

 

 

 

 

 

 

 

Operating Voltage Range

195-253 V~, 50 Hz

 

97-127 V~, 60 Hz

 

 

187-255 V~, 50 Hz

 

 

 

 

 

 

 

 

 

218-253 V~, 60 Hz

 

 

 

 

 

 

 

 

 

 

Oxygen Percentage

1-4 LPM = 93%±3%

 

 

 

 

1-5 LPM=93%±3%

 

 

 

 

 

 

 

 

 

 

 

Operating Altitude

 

 

 

 

 

 

 

 

 

(tested at 21°C only)

Across the voltage range:

 

Across the voltage range:

 

 

Across the voltage range:

 

0-1500 M (0-4921 ft)

No degradation of performance

 

No degradation of performance

 

No degradation in performance

 

 

 

 

Tested at nominal voltage only:

 

Tested at 230V/50Hz only:

 

1500-3000 M (4921-9842 ft)

Not recommended/Not tested

 

No degradation in performance

 

~4% oxygen perfomance drop at 5 LPM

 

3000-4000 M (9842-13123 ft)

Not recommended/ Not tested

 

Not recommended/Not tested

 

Not recommended/Not tested

 

 

 

 

 

 

 

 

 

 

Operating Environment Range*

No degradation in performance across

 

No degradation in performance across

 

~4% oxygen performance drop at 5 LPM

 

10°C to 35°C, humidity range of 30% to 75%

the operating voltage range

 

the operating voltage range

 

and 35°C condition. No degradation in

 

 

(tested at sea level).

 

(tested at sea level).

 

 

performance across the operating voltage

 

 

 

 

 

 

 

 

 

range at other liter flows (tested at 670 M).

 

 

 

 

 

 

 

 

 

 

 

Power Consumption

295 Watts Average

 

400 Watts Average

 

 

230v / 50 Hz - 365 Watts Average

 

 

 

 

 

 

 

 

 

230v / 60 Hz - 435 Watts Average

 

 

 

 

 

 

 

 

 

 

 

Weight

53 lbs. (24.5 Kilograms)

 

52 lbs. (23.5 Kilograms)

 

 

53 lbs. (24.5 Kilograms)

 

 

 

 

 

 

 

 

 

 

Sound Level (ISO 8359:1996 from front)

50.5 dbA overall average

 

52.5 dbA overall average**

 

50.5 dbA overall average (50Hz)

 

 

 

 

 

 

 

 

 

 

Dimensions

 

 

27.75"H x 16"W x 14"D (70.5 x 40.6 x 35.6 cm)

 

 

 

 

 

 

 

 

 

 

 

Pressure Relief Valve

 

 

44 psig±3psig (303 kPa±21 kPa)

 

 

 

 

 

 

 

 

 

 

 

Operating System

 

 

Time Cycle / Pressure Swing

 

 

 

 

 

 

 

 

 

For units equipped with an OSD,

Units with serial numbers less than H20000: 85%±2% (The audible alarm will alert at approximately 75%)

 

the visible “low oxygen” indicator

 

Units with serial numbers of H20000 and greater: 83.5%±1.5%

 

 

will activate at the following level

(The audible alarm will alert at approximately 75%. At less than 60%, the red “service required” light will activate.)

 

 

 

 

 

 

 

 

 

 

Storage Conditions

 

-40°C to 70°C, humidity range of 10% to 100%, including condensation

 

 

 

 

 

 

 

 

 

 

 

 

Equipment Class and Type

 

Class II Equipment Double Insulated

 

 

Type B Applied Part

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Approval Body

TUV

 

CSA

 

 

TUV

 

and

IEC 601-1+A1+A2

 

CAN/CSA-C22.2 No. 601.1-M90

 

IEC 601-1+A1+A2

 

Safety Standard

EN 60601+A1+A2

 

 

 

 

 

 

EN 60601+A1+A2

 

 

ISO 8359:1996

 

 

 

 

 

 

ISO 8359:1996

 

 

 

 

 

 

 

 

 

(515KZ excluded)

 

 

 

 

 

 

 

 

 

 

 

EMC Compliance To:

 

 

IEC 601-1-2

 

 

 

 

 

 

 

 

 

 

 

 

 

*Note: The OSD performance at 10°C to 35°C, 95% R.H. through voltage range on the 515DS verified at 670m.

**Note: This corresponds with a sound level of 50 dbA overall average as measured according to the guidelines of the now defunct ANSI Z79.13-1981 standard. Specifications subject to change without notice.