Appendix A: Performance Specification

Table A- 3: Waveform Horizontal Deflection

CATEGORY

Sweep

DESCRIPTION

Req: Synchronization: Sweep triggered by horizontal and vertical sync pulses

RI: Sweep Length: ≈12 divisions

RI: Sweep freeruns without input

Sweep Timing Accuracy

Req: 1 Line:

5 s/division ±1%

 

2 Line:

10 s/division ±1%

 

RI:

1 Field: displays one full field, including

 

 

field rate sync

 

 

2 Field: displays two full fields and

 

 

the field rate sync between them

 

 

Sweep Linearity

Req: 1 line: ±1%

 

2 line: ±1%

 

 

Magnified Sweep Accuracy

Req: 1 Line: 0.2 s/division ±1%

 

2 Line: 1.0 s/division ±1%

 

 

Magnified Sweep Linearity

Req: 1 line: ±1%

 

2 line: ±1%

 

 

Horizontal Position Range

Req: Any portion of the synchronized sweep can be

 

positioned on screen in all sweep modes.

 

 

External Horizontal Input

Req: 2 divisions/volt, ±2%

 

RI:

Menu is selected and enabled by RE-

 

 

MOTE connector ground closure.

 

 

 

Remote Sync

RI:

Input Amplitude: TTL level

 

RI:

Frequency: 25 Hz to 100 Hz positive

 

 

edge-triggered sweep

 

RI:

Enabling Signal: TTL low or ground

 

 

closure

 

 

 

 

 

 

 

 

A-6

1740A/1750A Series Waveform/Vector Monitor User Manual