Basic Measurements

Basic Measurements contains 1740A/1750A-Series graticule descriptions, followed by specific measurement procedures.

This monitor uses an internal graticule, which combines waveform and vector markings. The internal graticule scales are on the same plane as the CRT phosphor, eliminating parallax errors. Graticule illumination can be adjusted through the CRT menu (page 3-24) to provide optimum brightness for viewing or photographing displays.

Waveform Graticule

There are three versions of the waveform graticule, illustrated in Figure 4-1, Figure 4-2, and Figure 4-3:

HNTSC composite video graticule (1740A and 1750A only)HCCIR composite graticule (1741A and 1751A only)HDual standard graticule (1745A and 1755A only)

Horizontal Scale

The 0 IRE (0.3V) graticule line is divided into 12 major divi- sions. Each division represents a unit of time determined by the SWEEP and MAG settings. Line sweep rates are as fol- lows:

SWEEP

TIME/DIV

MAG ON

 

 

 

1 LINE SWEEP

5 µS/Div

200 nS/Div

2 LINE SWEEP

10 µS/Div

1 µS/Div

In 2 FIELD SWEEP with MAG ON, the entire vertical interval can be viewed.

1740A/1750A Series Waveform/Vector Monitor User Manual

4-1