Using the Menus

menu function is still in effect (LED is lighted), push the menu button to bring back the full display, then push it again to exit the menu.

Filter Menu (WAVEFORM Mode Only)

The FILTER menu is available for WAVEFORM mode only. Note that the instrument must be in WAVEFORM mode to access the R-Y displays. When the FILTER button is pushed in WAVEFORM mode, an on-screen menu allows the following selections. The bezel buttons are self-cancelling.

HFLAT — provides flat (normal) response.

HLUM — provides a low-pass response to display the luminance portion of the composite video signal.

HCHROM — provides bandpass response centered on the chrominance subcarrier frequency, and displays frequencies around the subcarrier frequency.

HDIFF — is a differentiated step filter (linearity steps are translated into impulses for amplitude compari- son).

HR--Yselects demodulated chrominance-versus- time display. The chrominance is demodulated on the R-Y (V) axis when burst is lined up on the normal axis. The VECTOR PHASE control can adjust the demodulator phase to any axis.

HSCH R--Y (1750A--Series Only) — demodulates the

SCH sync-locked oscillator. This display is useful for viewing SCH variations versus time.

HF + L — displays the video signal both flat and luminance filtered. The display is in parade or overlay mode, as selected through the on-screen menu. When PARADE is selected, the flat display is on the left.

1740A/1750A Series Waveform/Vector Monitor User Manual

3-7