Functional Check
16 P7225 2.5 GHz 10X Active Probe UserManua l
3. Using a ground lead and SMT KlipChip, connect the probe
ground to the instrument ground on the host instrument.
4. Using a standard tip, hold the probe to the center c onductor of the
PROBE COMPENSATION output on the host instrument.
5. Press AUTOSET, or adjust the host instrument to display a stable
calibration waveform.(You may need to adjust the probeoffset to
display the waveform.)
NOTE. Now is a good time to perform a probe calibration routine
from your instrument. Refer to the online help of your host instrument
for calibrationroutine procedures.
6. Disconnect the probe tip from the host instrument, a nd ground the
probe tip. (Connectthe KlipChip to the probe tip.)
7. Set the probe offset to 0.0 V to make the host instrument display
at the ground reference.
8. Adjustthe V/div to keep the waveform on the screen.