OperatingBasics
P7225 2.5 GHz 10X Active Probe UserManua l 21
1.7in Y lead adapter with 2 SMT KlipChips
Probewith only short ground pogo pin
Testpulse without probe
3in G round lead on square pin
Figure 5: Typicaleffects of using a probe tip adapter
Ground Lead LengthWhen you are probing a circuit, always use as short a ground lea d as
possible between the probe head and circuit ground.
The series inductance added by the probe tip a nd ground lead can
result in a resonantcircuit; this circuit may cause parasitic ringing
within the bandwidth of your host instrument. Refer to Figure 6.