Teledyne 2002 Interface Connector Pin Assignments for RS-232, Operation of the Serial Interface

Models: 2002

1 38
Download 38 pages 51.68 Kb
Page 24
Image 24

6.1.1.Interface Connector Pin Assignments for RS-485

(Full duplex-4 wire)(Jumper position 2-3).

PIN

MNEMONIC

SIGNAL

DESCRIPTION

 

 

 

 

2

TX+

Transmit +

Differential data signal levels to

8

TX-

Transmit -

the RS-485 bus

 

 

 

 

 

 

 

 

3

RX+

Receive +

Differential data signal levels to

4

RX-

Receive -

the RS-485 bus

6.1.2. For RS-485 (Half Duplex – 2 wire)

(Jumper position 1-2).

PIN

MNEMONIC

SIGNAL

DESCRIPTION

 

 

 

 

2

TX+/RX+

Transmit +/Receive+

Differential data signal levels to

8

TX-/RX-

Transmit -/Receive -

the RS-485 bus

6.2. Interface Connector Pin Assignments for RS-232

Note: an alternate method of connecting to a 2 wire bus is to leave the interface configured for full duplex and externally connect pins 2 and 3 to the (+) bus and pins 4 and 8 to the (-) bus.

6.3.Operation of the Serial Interface

Communication with the serial interface of the Model 2002 is via an ASCII data string. In the RS-232 mode the command message consist only of a command string and the terminator. The attention character and address string are not required, but if they are used they MUST be valid. If all components of the ASCII data string are valid the command will be accepted and executed. The RS-232 mode is sometimes referred to as point-to-point mode since only one device may be connected to the controller at any given time.

A message to the Model 2002 in the RS-485 mode consists of an attention character followed by the address string, the command string, and the terminator. If all components of the ASCII data string are valid the command will be accepted and executed. The RS-485 mode is also referred to as multipoint mode since up to 31 devices may be connected to the same controller in a network scheme.

The RS-485 address may be display on the model 2002 front panel (with software version 1.60 or higher). Press the MODE switch until the units field is lit, press both the UP nad DOWN switches simultaneously. The units light will now flash. Press the ZERO switch to display the RS-485 address (the address is displayed in it’s decimal form). Use the UP or DOWN switch to modify the address. Use the MODE switch to exit this function and save the new address into EEPROM memory.

COMMAND SYNTAX

In the following examples of syntax codes, the special characters are explained:

The characters in square brackets [ ] represents a command string, either upper or lower case command characters accepted. All characters must follow each other in the string with no spaces or other characters.

The characters within wavy brackets { } contain choices for the appropriate command.

The characters within the symbols < > are the common abbreviations for the one digit ASCII control codes which they represent, (e.g. <CR> represents carriage return).

When entering more than one command in the same data string, they must be separated by a comma.

All command strings must be followed by the terminator character (carriage return <CR>, also known as ENTER).

149-072011 Model 2002 Vacuum Gauge

Page 24 of 38