When a lower case character is present in an example it represents an option.
Character | Description | Valid Inputs: |
a | ........................................................01 - FF | |
m | Most Significant Digit Of Mantissa | 1 - 9 |
d | Decimal Digit | 0 - 9 |
e | Exponent | 0 - 5 |
u | Unit Of Pressure | T, M, or P |
, | Command Separator (comma) | N/A |
<CR> | Command Terminator (carriage return) | N/A |
Interrogation Commands
Command Description | Format | Sample Response | |
Transmit Averaged Pressure | P<CR> | Pa: 1.23456e+0 Torr<CR> | |
Transmit Pirani Pressure | R<CR> | Pr: | |
Transmit Piezo Pressure | Z<CR> | Pz: 7.65432e+2 Torr<CR> | |
Transmit | A<CR> | Multidrop Address: 01<CR> | |
Transmit Decimation Ratio | D<CR> | Decimation Ratio: 255<CR> | |
Transmit Selected Gas # | G<CR> | Gas#: 0<CR> | |
Transmit High Set point | H<CR> | Hi: 1.00000e+1 Torr<CR> | |
Transmit Low Set point | L<CR> | Lo: | |
Transmit Device Status | S<CR> | 00044<CR> | |
Transmit Turnaround Delay | T<CR> | Comm Delay: 6<CR> | |
Transmit Selected Units | U<CR> | Torr<CR> | |
Transmit Software Version# | V<CR> | Hastings Instruments - Model 2002 | |
| .................................................. | Version 1.60 - | |
Parameter Modification Commands |
|
| |
Command Description | Format | Valid Range: | |
Modify High Set point | H={m.dd}E{+e}<CR> | ||
Modify Low Set point | |||
Modify Selected Gas # | G={d}<CR> | 0 To 4 | (Decimal) |
Modify Selected Units | U={u}<CR> | T, or M, ............{or P, if Available} | |
Modify Decimation Ratio | D={dddd}<CR> | 63 To 7936 | (Decimal) |
Modify | *{aa}A={aa}<CR> | 1 To FF | (Hexadecimal) |
Modify Turnaround Delay | *{aa}T={dd}<CR> | 0 To 255 | (Decimal) |
Page 25 of 38 |