Page 25
Image 25

When a lower case character is present in an example it represents an option.

Character

Description

Valid Inputs:

a

RS-485 Address (hexadecimal 0-9, A-F)

........................................................01 - FF

m

Most Significant Digit Of Mantissa

1 - 9

d

Decimal Digit

0 - 9

e

Exponent

0 - 5

u

Unit Of Pressure

T, M, or P

,

Command Separator (comma)

N/A

<CR>

Command Terminator (carriage return)

N/A

Interrogation Commands

Command Description

Format

Sample Response

Transmit Averaged Pressure

P<CR>

Pa: 1.23456e+0 Torr<CR>

Transmit Pirani Pressure

R<CR>

Pr: 1.98765e-3 Torr<CR>

Transmit Piezo Pressure

Z<CR>

Pz: 7.65432e+2 Torr<CR>

Transmit RS-485 Address

A<CR>

Multidrop Address: 01<CR>

Transmit Decimation Ratio

D<CR>

Decimation Ratio: 255<CR>

Transmit Selected Gas #

G<CR>

Gas#: 0<CR>

Transmit High Set point

H<CR>

Hi: 1.00000e+1 Torr<CR>

Transmit Low Set point

L<CR>

Lo: 1.00000e-2 Torr<CR>

Transmit Device Status

S<CR>

00044<CR>

Transmit Turnaround Delay

T<CR>

Comm Delay: 6<CR>

Transmit Selected Units

U<CR>

Torr<CR>

Transmit Software Version#

V<CR>

Hastings Instruments - Model 2002

 

..................................................

Version 1.60 - (07-02-2002)<CR>

Parameter Modification Commands

 

 

Command Description

Format

Valid Range:

Modify High Set point

H={m.dd}E{+e}<CR>

1.00000e-9 to 9.99999e+9

Modify Low Set point

L={m.dd}E{-e}<CR>

1.00000e-9 to 9.99999e+9

Modify Selected Gas #

G={d}<CR>

0 To 4

(Decimal)

Modify Selected Units

U={u}<CR>

T, or M, ............{or P, if Available}

Modify Decimation Ratio

D={dddd}<CR>

63 To 7936

(Decimal)

Modify RS-485 Address

*{aa}A={aa}<CR>

1 To FF

(Hexadecimal)

Modify Turnaround Delay

*{aa}T={dd}<CR>

0 To 255

(Decimal)