Reset / Restore Commands

Reset / Restore Commands

Command Description

Format

Notes:

Escape

<Esc>

Reset Command Buffer

 

.......................................................................................

(Ignore prior Input)

Software Reset

/R<CR>

Reinitialize Software

Restore Factory Defaults

/#<CR>

Restore Calibration Register Default Values

Set Zero

/0 <CR>

Store Present Pressure as Instrument

 

..........................................

ZERO (if Piezo <32 Torr & Pirani <50 mTorr)

Force Zero

/!<CR>

Override Limit Checks And Store Present

 

.......................................................................

Pressure as Instrument ZERO

Device Status

When requested to transmit its status the Model 2002 responds with a five digit number which is explained in the following:

Digit #

1

2

3

 

4

5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Serial Receiver Overflow

= 4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Main Board EEPROM Error

= 2

 

 

 

 

 

Probe EEPROM (any) Error

= 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Probe EEPROM Not Responding

 

 

 

 

 

 

 

= 8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Probe EEPROM Read Error

 

= 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Probe EEPROM Checksum Error

 

= 4

 

 

 

Probe EEPROM Verification Error

 

= 2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Probe EEPROM Identification Error

 

= 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Communications Syntax Error

 

 

 

 

= 4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Piezo Sensor Bad (voltage out of range)

 

 

 

= 2

 

Pirani Sensor Bad (voltage out of range)

 

 

 

= 1

 

 

 

 

 

 

 

 

 

 

 

 

 

High Setpoint Alarm (pressure exceeds setpoint)

 

 

 

 

 

 

 

 

 

 

 

= 4

Low Setpoint Alarm (pressure less than setpoint)

 

 

 

= 2

Gas# Changed (not the same as when unit last calibrated)

 

= 1

149-072011 Model 2002 Vacuum Gauge