Teledyne 2002 instruction manual Specifications

Models: 2002

1 38
Download 38 pages 51.68 Kb
Page 5
Image 5

board is also available. The Model 2002 instrument accepts an external TTL level "Remote Zero" to zero the instrument when another instrument (e.g. an ion gauge) indicates that the pressure is less than 1x10-5 Torr. The standard analog output is linear per decade directly corresponding to the control unit’s display.

The Model 2002 comes calibrated for nitrogen. Conversion factors for other gases are selectable by the user (see section 3.5). This direct readout display allows for flexibility in user system configurations which operate with various gases.

An optional serial interface board can be installed to provide RS232 or RS485 support. The Model 2002 control unit fits a standard DIN 43700 insert (1/4 DIN) and mounts into a 3.62 in. (92 mm) square hole. The case is constructed of NORYL rated UL-94-V1 (self-extinguishing plastic) and is equipped with mounting brackets that are adjustable up to 3/16” (4.8mm).

1.4. Specifications

Measuring range

1x10-4to 10+3 Torr

....................................................................................................................................

1.3x10-4to 1.3x10+3 mbar

Accuracy (N2, T=230C)

+ 20% of reading (5x10-4to 5x10+1 Torr)

.........................................................................................................

+ 1.5% of reading (5x10+1 to 1x10+3 Torr)

Ambient temperature operating range

00 to 500C

Bakeout temperature (with transducer electronics removed)

2500C

Analog output

1.0-4.5 Volts (0.5 VDC linear per decade)

Process control

2 TTL outputs (1 TTL auto zero input)

Digital readout

four green LED - resolution is three digit (plus exponent)

Equipment operating range

120/240 V~, 0.25A, 50/60 Hz

.........................................................................................

A 6 ft., 3-wire power cord is provided with each unit.

Fuse

250 V, 1/4 A.

Transducer mounting

Any position without recalibration.

Transducer internal volume

<1.5cc

Wetted material

Au, Si3N4, Si, Pt, PyrexTM,

............................................................................................................................................

KovarTM and 316 SS

Stability with voltage

Variation in reading undetectable

...................................................................................................

as power fluctuates within power specification.

Weight (display, cable, and sensor)

Net:

2.5 lbs., Shipping: 3.4 lbs.

Transducer response time

typically, less than 150 msec for a step

....................................................................................................................

change from 10-3Torr to 10+3 Torr.

Positive pressure

150 psig, Recalibration may be required

.....................................................................

if unit is exposed to pressures greatly exceeding measuring range.

149-072011 Model 2002 Vacuum Gauge