board is also available. The Model 2002 instrument accepts an external TTL level "Remote Zero" to zero the instrument when another instrument (e.g. an ion gauge) indicates that the pressure is less than
The Model 2002 comes calibrated for nitrogen. Conversion factors for other gases are selectable by the user (see section 3.5). This direct readout display allows for flexibility in user system configurations which operate with various gases.
An optional serial interface board can be installed to provide RS232 or RS485 support. The Model 2002 control unit fits a standard DIN 43700 insert (1/4 DIN) and mounts into a 3.62 in. (92 mm) square hole. The case is constructed of NORYL rated
1.4. Specifications
Measuring range | ||
.................................................................................................................................... | ||
Accuracy (N2, T=230C) | + 20% of reading | |
......................................................................................................... | + 1.5% of reading (5x10+1 to 1x10+3 Torr) | |
Ambient temperature operating range | 00 to 500C | |
Bakeout temperature (with transducer electronics removed) | 2500C | |
Analog output | ||
Process control | 2 TTL outputs (1 TTL auto zero input) | |
Digital readout | four green LED - resolution is three digit (plus exponent) | |
Equipment operating range | 120/240 V~, 0.25A, 50/60 Hz | |
......................................................................................... | A 6 ft., | |
Fuse | 250 V, 1/4 A. | |
Transducer mounting | Any position without recalibration. | |
Transducer internal volume | <1.5cc | |
Wetted material | Au, Si3N4, Si, Pt, PyrexTM, | |
............................................................................................................................................ | KovarTM and 316 SS | |
Stability with voltage | Variation in reading undetectable | |
................................................................................................... | as power fluctuates within power specification. | |
Weight (display, cable, and sensor) | Net: | 2.5 lbs., Shipping: 3.4 lbs. |
Transducer response time | typically, less than 150 msec for a step | |
.................................................................................................................... | change from | |
Positive pressure | 150 psig, Recalibration may be required | |
..................................................................... | if unit is exposed to pressures greatly exceeding measuring range. |
Page 5 of 38 |