8.2.Operation

8.2.Operation

The 4-20 mA output option board provides dual channel current output linearly proportional to the HPM 2002’s pressure reading. The first channel corresponds to the higher pressure range (0-1024 Torr). The second channel corresponds to the lower pressure range (0-1000mTorr). The equations below give the output currents for a given pressure:

16mA

I (channel1) = 4mA + (P(Torr))* ⎜

 

 

1024Torr

16mA

 

I (channel2) = 4mA + (P(mTorr))* ⎜

 

 

 

 

1000mTorr

Note that for both channels, the output is always between 4mA and 20mA. Specifically, when the pressure is below 1 Torr, channel 1 current will be approaching 4mA. Also, when the pressure is above 1 Torr channel 2 current will be saturated at 20mA.