Teledyne 2002 instruction manual Communications Option Board, RS-232-E Interface Specifications

Models: 2002

1 38
Download 38 pages 51.68 Kb
Page 23
Image 23
6. Communications Option Board

6. Communications Option Board

Both RS-232 and RS-485 communication interface boards are available as an option for the Model 2002. The communication option boards allow data to be output to a host computer with the appropriate interface. Connection to the RS-232 or RS-485 communication interface boards are via a 9 pin D-subminiature female connector.

6.1.RS-232-E Interface Specifications

Format

EIA standard RS-232-E, full-duplex, no handshaking, asynchronous

Data Rate

9600 baud

Character Length

Eight data bits

Parity

No parity

Stop bits

One stop bit

A common application of the RS232 version of the HPM-2002, is to connect the pressure gauge directly to the serial port of a PC. This is done by first wiring a communication cable in the manner shown below.

PIN

MNEMONIC

SIGNAL

DESCRIPTION

 

 

 

 

2

TX

Transmit Data

Transmits data within RS-232-E voltage levels

3

RX

Receive Data

Accepts data within RS-232-E voltage levels

5

SG

Signal Ground

Establishes the common ground reference

potential for all interchange circuits

when

 

 

Communicating via RS-232-e

232-E.

 

 

 

 

 

 

 

 

Host computer ready to accept data.

4

CTS

Clear to Send

(Reserved for future use, not implemented

 

 

 

by present software.

8

RDY

Ready

Ready to receive data (always high,

not implemented by present software).

 

 

 

149-072011 Model 2002 Vacuum Gauge

Page 23 of 38

Page 23
Image 23
Teledyne 2002 instruction manual Communications Option Board, RS-232-E Interface Specifications