Installation 3000ZA
Teledyne Analytical Instruments 18
If greater sample flow is required for improved response time,
install a bypass in the sampling system upstream of the analyzer input.
EXHAUST OUT:
Exhaust connections must be consistent with the hazard level of the
constituent gases. Check local, state, and federal laws, and ensure that
the exhaust stream vents to an appropriately controlled area, if required.
ZERO IN and SPAN IN (Optional):
These are additional ports for inputting span and zero gas. There are
electrically operated valves inside for automatic switching between
calibration and sample gases. These valves are completely under control
of the 3000ZA electronics. They can be externally controlled only
indirectly through the remote cal inputs as described below.
Pressure, flow, and safety considerations are the same as prescribed for
the SAMPLE IN inlet above.
3.3.2 Electrical Connections
For safe connections, no uninsulated wiring should be able to come
in contact with fingers, tools or clothing during normal operation.
CAUTION: USE SHIELDED CABLES. ALSO, USE PLUGS THAT
PROVIDE EXCELLENT EMI/RFI PROTECTION. THE
PLUG CASE MUST BE CONNECTED TO THE CABLE
SHIELD, AND IT MUST BE TIGHTLY FASTENED TO
THE ANALYZER WITH ITS FASTENING SCREWS.
ULTIMATELY, IT IS THE INSTALLER WHO ENSURES
THAT THE CONNECTIONS PROVIDE ADEQUATE
EMI/RFI SIELDING.
3.3.2.1 Primary Input Power
The power cord receptacle and fuse block are located in the same
assembly. Insert the power cord into the power cord receptacle.
CAUTION: POWER IS APPLIED TO THE INSTRUMENT'S
CIRCUITRY AS LONG AS THE INSTRUMENT IS
CONNECTED TO THE POWER SOURCE. THE RED
SWITCH ON THE FRONT PANEL IS FOR SWITCHING
POWER ON OR OFF TO THE DISPLAYS AND
OUTPUTS ONLY.
The universal power supply requires a n 85–250 VAC, 47-63 Hz
power source.