Trace Oxygen Analyzer | Operation | |
|
|
|
4.5 The Alarms Function
The Model 3000ZA is equipped with 2 fully adjustable concentration alarms and a system failure alarm. Each alarm has a relay with a set of form “C" contacts rated for 3 amperes resistive load at 250 VAC. See Table
The system failure alarm has a fixed configuration as described in chapter 3 Installation.
The concentration alarms can be configured from the front panel as either high or low alarms by the operator. The alarm modes can be set as latching or
Decide how your alarms should be configured. The choice will depend upon your process. Consider the following four points:
1.Which if any of the alarms are to be high alarms and which if any are to be low alarms?
2.Setting an alarm as HIGH triggers the alarm when the oxygen concentration rises above the setpoint. Setting an alarm as LOW triggers the alarm when the oxygen concentration falls below the setpoint.
3.Decide whether you want the alarms to be set as:
•Both high (high and
•One high and one low alarm, or
•Both low (low and
4.Are either or both of the alarms to be configured as failsafe?
5.In failsafe mode, the alarm relay
6.Are either of the alarms to be latching?
7.In latching mode, once the alarm or alarms trigger, they will remain in the alarm mode even if process conditions revert