Trace Oxygen Analyzer

Installation

 

 

 

pressure applications (5-50 psig) use the restrictor that has a blue sticker on the body.

Pressurized service: Connect the restrictor with the blue dot to the Sample In port. The small circular orifice should face AWAY from the back of the unit (against the direction of gas flow).

Note: Remove the blue sticker from the restrictor before using.

Low pressure applications: Connect the restrictor without the blue sticker in the same manner (less than 5 psig).

Vacuum Service: For applications involving pressure less than 5-10 in Hg) connect the restrictor without the blue sticker to the Exhaust Out port. The small circular orifice should face TOWARD the back of the unit (against the direction of gas flow).

CAUTION: OPERATING THE UNIT WITHOUT RESTRICTORS CAN

CAUSE DAMAGE TO THE SENSOR.

The unit is manufactured with 1/4” tube fittings and includes 6 mm adapters for metric system installations. For a safe connection:

1.Insert the tube into the fitting and finger tighten the nut until the tubing cannot be rotated freely by hand in the fitting. This may require an additional 1/8 turn beyond finger tight.

2.Hold the fitting body steady with a backup wrench. using another wrench, rotate the nut another 1-1/4 turns.

SAMPLE IN:

In the standard model, gas connections are made at the SAMPLE IN and EXHAUST OUT connections. Calibration gases must be tee'd into the Sample inlet with appropriate valves.

The inlet gas pressure should be reasonably regulated. Pressures between 2 and 50 psig are acceptable as long as the pressure, once established, will keep the front panel flowmeter reading in an acceptable range (0.5 to 2.0 SLPM). For non-pressurized sample or very low pressure, (less than 2 psig) vacuum service plumbing is recommended. Exact figures will depend on your process.

Teledyne Analytical Instruments

17

Page 29
Image 29
Teledyne 3000ZA operating instructions Sample