Trace Oxygen Analyzer

Appendix

 

 

 

 

 

 

Power: Universal power supply 85-250 V ac, at 47-63 Hz.

Operating Temperature: 5-35 °C

Accuracy: ±2% of full scale at constant temperature (at ca;obrated ramge).

±5% of full scale over operating temperature range, except 0-1 ppm analysis range, once thermal equilibrium is reached.

±0.2 ppm on 0-1 ppm analysis range, once thermal equilibrium is reached.

Analog outputs: 0-1 VDC percent-of-range, 0-1 VDC range ID.

4-20 mA DC (isolated) percent-of-range,

4-20 mA DC (isolated) range ID (optional).

Dimensions: 19 cm high, 24.9 cm wide, 31 cm deep (6.96 in high, 8.7 in wide, 12.2 in deep).

Teledyne Analytical Instruments

Page 71
Image 71
Teledyne 3000ZA operating instructions Trace Oxygen Analyzer Appendix