Trace Oxygen Analyzer

Operation

 

 

 

If an alarm is tripped, the second line will change to show which alarm it is:

0.0ppm Anlz

AL—1

Note: If you log off the system using the logout function in the system menu, you will now be required to re-enter the password to gain access to Span, Zero, Alarm, and Range functions.

4.3.4Logout

The Logout function provides a convenient means of leaving the analyzer in a password protected mode without having to shut the instrument off. By entering Logout, you effectively log off the instrument leaving the system protected against use until the password is reentered. To log out, press the System button to enter the System function.

TRAK/HLD Auto—Cal

PSWD Logout More

Use the __ arrow keys to position the blinking over the Logout function, and press Enter to Log out. The screen will display the message:

Protected Until

Password Reentered

4.3.5System Self-Diagnostic Test

The Model 3000ZA has a built-in self-diagnostic testing routine. Pre-programmed signals are sent through the power supply, output board and sensor circuit. The return signal is analyzed, and at the end of the test the status of each function is displayed on the screen, either as OK or as a number between 1 and 3. (See System Self Diagnostic Test in Chapter 5 for number code.)

The self diagnostics are run automatically by the analyzer whenever the instrument is turned on, but the test can also be run by the operator at any time. To initiate a self diagnostic test during operation:

Press the System button to start the System function.

Teledyne Analytical Instruments

37

Page 49