Percent Oxygen Analyzer

Operational Theory 2

 

 

 

 

 

 

 

 

 

2.3Electronics

2.3.1 General

The signal processing uses an Intelmicrocontroller with on-board RAM and ROM to control all signal processing, input/output, and display functions for the analyzer. System power is supplied from a universal power supply module designed to be compatible with most international power sources.

The power supply circuitry is on the Power Supply PCB, which is mounted vertically, just behind the rear panel of the Control Unit.

The signal processing electronics including the sensor amplifier, microcontroller, analog to digital, and digital to analog converters are located on the Main PCB, which is mounted vertically, just behind the front panel of the Control Unit.

2.3.2 Signal Processing

Figure 2-3 is a block diagram of the signal processing electronics described below.

 

 

 

Millivolt

 

 

 

Potput

 

 

TEM PERATU RE

SENSO R

M FC

 

CO M PENSATION

 

NETW ORK ON

AM PLIFIER

 

 

Microamp

SENSO R PCB

 

 

 

Output

 

 

 

 

I-22 Sensor

 

KEYBO ARD

A D C

M ICRO - CO NTRO L- LER

E –I CON V

 

4–20 m A dc

 

Concentration

0–10 V dc

 

D A C

 

RAN GE ID

 

 

 

 

 

 

 

 

D IS P LAY

 

 

A LAR M S

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RELAYS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RELAYS

 

 

A LAR M S

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 2-3: Block Diagram of the Signal Processing Electronics

Teledyne Analytical Instruments

2-5

Page 15
Image 15
Teledyne 3300PA operating instructions Electronics, General, Signal Processing