4 Operation | Model 3300PA | |
|
|
|
|
|
|
|
|
|
Note: When you stop pressing the UP/DOWN keys, the rapid flashing will cease and the new span value will be acquired by the analyzer after five seconds.
The alarms will only be nonresponsive for 60 seconds. This time- frame allows you to reintroduce sample gas into the analyzer.
5. Immediately flow sample into the analyzer.
Supplementary Information
If, during the Span Procedure, you pressed the SPAN button by mis- take, you must wait five minutes for the analyzer to resume analisis or you can press the UP button and then the DOWN button. (Pressing the UP and DOWN buttons causes the analyzer to
If during the span procedure, you press the RANGE or ALARM buttons:
-either the range or alarm routine will be activated.
-any changes to span will be rejected.
-the 60 second alarm delay will not occur, i.e., the alarms will be responsive immediately.
CAUTION: TET/AI considers the action of pressing the ALARM or RANGE buttons to be at your (the user’s) discretion and NOT under guarantee of alarm protection.