SCPI Command Reference
C–18 TM-XDOP-01XN
Table C-7Commands for Fold Protection
Function SCPI Command Description Query
Set Output Fold
Delay [:]OUTPut[<channel>]:PROTection:FOLD:DELay
<delay> Set the delay time (seconds)
before fold protection is triggered. Yes
Set Output Fold
Mode [:]OUTPut[<channel>]:PROTection:FOLD[:MODE]
{NONE|CC|CP|CV} Select which regulation mode to
fold back (None, CV, CC, CP) Yes
Query Fold
Protection Tripped [:]OUTPut[<channel>]:PROTection:FOLD:TRIPped? Query if fold protection has
tripped N/A
Table C-8Commands for Triggering
Function SCPI Command Description Query
Set Immediate
Initiation of Trigger
System
[:]INITiate[<channel>][:IMMediate] Initiate a triggered event or
sequence N/A
Set Trigger Source [:]TRIGger[<channel>][:SEQuence]:SOURce
{BUS|EXTernal|IMMediate|NONE} Sets the trigger source for
triggered setpoints Yes