Emerson 396PVP instruction manual Ordering Information, Percent Linearity

Page 10

Section 1: Description and Specifications

396P + 396PVP Sensors Instruction Manual

August 2013

LIQ_MAN_396P_396PVP

 

 

Integral Cable 396P: Code 01 - 25 ft; Code 02 - 15 ft coaxial / 396PVP: none - must use mating VP cable

Preamplifier Options: Remote or Integral SMART preamplifier

Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/2 lb)

Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/

*Percent Linearity

pH Range

396 / 396VP

396P / 396PVP

396R / 396RVP

 

 

 

 

 

 

 

 

GPHT Hemi

GPHT Flat

GPHT Hemi

GPLR Flat

GPHT Hemi

GPHT Flat

 

 

 

 

 

 

 

0-2 pH

94%

95%

94%

94%

93%

 

 

 

 

 

 

 

2-12 pH

99%

99%

97%

98%

97%

98%

 

 

 

 

 

 

 

12-13 pH

97%

96%

98%

95%

98%

95%

 

 

 

 

 

 

 

13-14 pH

92%

98%

98%

 

 

 

 

 

 

 

1.3Ordering Information

The 396P Sensor is housed in a molded reinforced polypropy- lene body with 1 in. MNPT threads suitable for insertion, submersion or flow through installation. The sensor includes a general purpose pH electrode or a platinum ORP electrode, a reference junction and a solution ground. The 396P comes standard with a recessed electrode; an optional slotted tip is also available. In addition, the 396P features an optional inte- gral hermetically sealed preamplifier and 15 ft or 25 ft cable lengths. Automatic temperature compensation, Pt 100 or 3K Balco, is standard with the 396P.

The 396P insertion/submersion sensor shown with the fully recessed bulb (standard option).

396P TUpH INSERTION/SUBMERSION POLYPROPYLENE pH/ORP SENSOR

CODE

PREAMPLIFIER/CABLE (Required Selection)

01

With integral SMART preamplifier, 25 ft (7.6m) cable

 

 

02

Without integral preamplifier, 15 ft (4.6m) cable

 

 

CODE

MEASURING ELECTRODE TYPE (Required Selection)

10

GPLR hemi bulb, General Purpose Low Resistivity (0-14 pH)

12ORP

13GPLR flat bulb, General Purpose Low Resistivity (2 - 13 pH)

CODE

ANALYZER/TC COMPATIBILITY (Required Selection)

50

For 1181 (3K TC)

54For 1054A/B, 2054, 2081 (Pt 100 RTD)

55For 54e, 3081, 81, 4081, 1055, 5081, Xmt (PT-100 RTD)

CODE

OPTIONAL SELECTION

 

41

Slotted Tip (not available on flat bulb sensors)

 

396P -

01 - 10 - 55

EXAMPLE

 

 

 

NOTE: The 396P is also compatible with Model SCL-P/Q (option 02-54 only).

4

Description and Specifications

Image 10
Contents TUpH Combination pH/ORP Sensors This page left blank intentionally Essential Instructions About This Document Contents Table of Contents Description and Specifications Features and ApplicationsDescription and Specifications Specifications General Available pH Accuglass Types Gplr hemi or flat glassOrdering Information Percent LinearityExamples of all sensing tip offerings Accessories Extension Cables required when using a remote junction boxUnpacking and Inspection InstallationMounting Installation Flow Through and Insertion MountingFlow-Through Tee with Adapter PN 40396P02 40396P03 Spray Wash Assembly PN Wiring GeneralWiring Wiring for 396P-01 Gray CableWiring for 396PVP Gray Cable and 54e pH/ORP Wiring for 396PVP 11. Wiring for Dual 396P-02 13. Wiring for 396P-01 Gray Cable and 1056/56 17. Wiring for 396PVP Gray Cable 20. Wiring for 396P-01 Gray Cable 3Wiring 27. Wiring for 396P-01 Blue Cable 29. Wiring for 396P-02 Blue Cable 31. Wiring for 396PVP Blue Cable 33. Wiring for 396P-01 Gray Cable and 5081-P-HT 37. Wiring for 396PVP Gray Cable and 5081-P-HT 41. Wiring for 396P-01 Gray Cable 43. Wiring for 396P-02 Gray Cable 45. Wiring for 396PVP Gray Cable 47. Wiring for 396P-01 Gray Cable and Xmt 51. Wiring for 396PVP Gray Cable and Xmt Wiring Wiring This page left blank intentionally Start-Up and Calibration 396P and 396PVP pH Sensors396P and 396PVP ORP Sensors Start-Up & CalibrationGeneral Information MaintenanceAutomatic Temperature Compensator 396P & 396PVP pH SensorsMaintenance 396P and 396PVP ORPCleaning Platinum Electrode This page left blank intentionally Diagnostics and Troubleshooting Troubleshooting with Advanced DiagnosticsTroubleshooting without Advanced Diagnostics Diagnostics & TroubleshootingDescription Diagnostics & Troubleshooting Warranty This page left blank intentionally Non-Warranty Repair Warranty RepairReturn of Material Emerson Process Management