Section 6: Diagnostics & Troubleshooting | 396P + 396PVP Sensors Instruction Manual |
August 2013 | LIQ_MAN_396P_396PVP |
|
|
6.2Troubleshooting without Advanced Diagnostics
Table
TABLE
Problem | Probable Cause | Remedy |
|
|
|
Meter reads off scale. (Display | Defective preamplifier | Replace preamplifier (for code 02 |
reads overrange). |
| sensors). For code 01, replace sensor. |
|
|
|
| T.C. element shorted | Check T.C. element as instructed |
|
| in Section 6.1 and |
|
| replace sensor if defective. |
|
|
|
| Sensor not in process. Sample | Make sure sensor is in process with |
| stream is low or air bubbles are | sufficient sample stream (refer to |
| present. | Section 2.0 for installation details). |
|
|
|
| Open glass electrode | Replace sensor. |
|
|
|
| Reference element open - no contact | Replace sensor. |
|
|
|
Display reads between 3 and 6 pH | Electrode cracked | Replace sensor. |
regardless of actual pH of solution |
|
|
or sample. |
|
|
|
|
|
Meter or display indication swings | T.C. element shorted | Check T.C. element as instructed |
or jumps widely in AUTO T.C. Mode. |
| in Section 6.1 and replace |
|
| sensor if defective. |
|
|
|
Span between buffers extremely | T.C. element open | Check T.C. element as instructed |
short in AUTO T.C. Mode. |
| in Section 6.1 and replace sensor |
|
| if defective. |
|
|
|
Sluggish or slow meter indication | Electrode coated | Clean sensor as instructed in |
for real changes in pH level. |
| Sections 6.2 or Section 6.3.2. |
|
| Replace sensor if cracked. |
|
|
|
| Electrode defective | Replace sensor. |
|
|
|
Transmitter cannot be standardized. | Electrode coated or cracked | Clean Sensor as instructed in |
|
| Sections 6.2 or Section 6.3.2 |
|
| Replace sensor if cracked. |
|
|
|
| Defective preamplifier | Replace preamplifier. |
|
|
|
Transmitter short spans between | Aged glass electrode or high | Replace sensor. |
two different buffer values. | temperature exposure |
|
|
|
|
| Electrode coated | Clean Sensor as instructed in |
|
| Section 6.2 or Section 6.3.2. |
|
| Replace sensor if cracked. |
|
|
|
| Air bubbles trapped in sensor end | Shake the sensor in solution. See |
| between glass bulb and sensor body | Section 2.0 for mounting guidelines. |
|
|
|
42 | Diagnostics and Troubleshooting |