Emerson 396PVP instruction manual Wiring, General

Page 19

396P + 396PVP Sensors Instruction Manual

Section 3: Wiring

LIQ_MAN_396P_396PVP

August 2013

 

 

Section 3: Wiring

3.1General

Figures in this section provide the guidelines for wiring the 396P-01 sensor to various

Analyzer/Transmitter instruments.

To determine which wiring guideline to use, locate the model number of the sensor to be installed.

1.If the cable needs to be extended, use a high quality eleven conductor double shielded instru- ment cable available from Rosemount Analytical. Refer to Table 3-1 for the appropriate junc- tion box to use and the corresponding wiring details.

NOTE: If the cable is too long, loop up the excess cable. If the cable has to be shortened, cut and terminate each conductor neatly and make sure that the overall (outermost) drain wire is not shorted out with either of the two inner drain wires (shields).

2.Signal cable should be run in a dedicated conduit (preferably an earth grounded metallic conduit) and should be kept away from AC power lines. For your convenience, a wire nut kit is furnished (in a plastic bag wrapped around the cable).

NOTE: For maximum EMI/RFI protection when wiring from the sensor to the junction box, the outer braid of the sensor should be connected to the outer braided shield of the extension cable. The outer braid of the extension cable to the instrument must be terminated at earth ground or by using an appropriate metal cable gland fitting that provides a secure connection to the instru- ment cable.

Wiring

The 396P and 396PVP has an optional built-in preamplifier and is offered with a shielded cable. The cable should be handled carefully and kept dry and free of corrosive chemicals at all times. Extreme care should be used to prevent it from being twisted, damaged or scraped by rough, sharp edges or surfaces.

DANGER

DO NOT CONNECT SENSOR CABLE TO POWER LINES. SERIOUS INJURY MAY RESULT.

NOTE: Remove electrical tape or shrink sleeve from gray reference wire before connecting wire to terminal.

Wiring

13

Image 19
Contents TUpH Combination pH/ORP Sensors This page left blank intentionally Essential Instructions About This Document Contents Table of Contents Features and Applications Description and SpecificationsDescription and Specifications Available pH Accuglass Types Gplr hemi or flat glass Specifications GeneralPercent Linearity Ordering InformationExamples of all sensing tip offerings Extension Cables required when using a remote junction box AccessoriesUnpacking and Inspection InstallationMounting Flow Through and Insertion Mounting InstallationFlow-Through Tee with Adapter PN 40396P02 40396P03 Spray Wash Assembly PN General WiringWiring for 396P-01 Gray Cable WiringWiring for 396PVP Gray Cable and 54e pH/ORP Wiring for 396PVP 11. Wiring for Dual 396P-02 13. Wiring for 396P-01 Gray Cable and 1056/56 17. Wiring for 396PVP Gray Cable 20. Wiring for 396P-01 Gray Cable 3Wiring 27. Wiring for 396P-01 Blue Cable 29. Wiring for 396P-02 Blue Cable 31. Wiring for 396PVP Blue Cable 33. Wiring for 396P-01 Gray Cable and 5081-P-HT 37. Wiring for 396PVP Gray Cable and 5081-P-HT 41. Wiring for 396P-01 Gray Cable 43. Wiring for 396P-02 Gray Cable 45. Wiring for 396PVP Gray Cable 47. Wiring for 396P-01 Gray Cable and Xmt 51. Wiring for 396PVP Gray Cable and Xmt Wiring Wiring This page left blank intentionally 396P and 396PVP pH Sensors Start-Up and CalibrationStart-Up & Calibration 396P and 396PVP ORP Sensors396P & 396PVP pH Sensors MaintenanceAutomatic Temperature Compensator General Information396P and 396PVP ORP MaintenanceCleaning Platinum Electrode This page left blank intentionally Troubleshooting with Advanced Diagnostics Diagnostics and TroubleshootingDiagnostics & Troubleshooting Troubleshooting without Advanced DiagnosticsDescription Diagnostics & Troubleshooting Warranty This page left blank intentionally Non-Warranty Repair Warranty RepairReturn of Material Emerson Process Management