396P + 396PVP Sensors Instruction Manual | Section 1: Description and Specifications |
LIQ_MAN_396P_396PVP | August 2013 |
|
|
The 396P and 396PVP sensors are available in two configurations: the standard shrouded tip, in which the pH glass is completely recessed within the reference junction for abrasive or rough applications, and the optional slotted tip with a partially exposed pH glass for viscous or low flow applications.
The 396P and 396PVP sensors feature a titanium solution ground, constructed in an annular design around the pH/ORP electrode. The solution ground provides advanced sensor diagnostics for preventative maintenance when used with Rosemount Analytical 54e, 56, 1056, 1057, 5081, 6081, 1066, and XMT instruments. In addition, the Model 396P sensor can be used with most
The 396P and 396PVP sensor is housed in a molded polypropylene body with EPDM seals, making it virtually indestructible and chemically resistant. Complete encapsulation eliminates leakage or high humidity problems traditionally found in other pH/ORP designs. The simplified construction, designed with user convenience in mind, does not require electrolyte (KCl) replen- ishment or any high maintenance troubleshooting procedures.
The 396 and 396VP TUpH Sensors are housed in a stainless steel body with optional 1" threaded connector suitable for insertion, submersion or flow through installation. The sensor includes a general purpose glass pH electrode in the standard bulb or optional flat glass configuration. The TUpH reference has a large area polypropylene junction with a gel filled reference electrolyte. The 396 and 396VP TUpH are available with an optional SMART integral preamplifier when choosing the
The 396R and 396RVP Sensors are designed for use with a
The Rosemount Analytical 54e, 56, 1056, 1057, 1066, 5081, 6081, and XMT instruments offer advanced sensor diagnostics to provide preventative maintenance by notifying the operator of the need for replacement and cleaning of an aged or fouled sensor, thus allowing continuous optimum performance. The 396R and 396RVP sensors also feature a slotted tip for protection from breakage while allowing the process to flow by the glass electrode for accurate and reliable pH measurement.
1.2Specifications - General
Measurements and Ranges: pH*:
Available pH ACCUGLASS Types: GPLR hemi or flat glass
Wetted Materials: Titanium, Polypropylene, EPDM, glass; platinum (ORP only)
Process Connection: 1 in. MNPT front and rear facing threads
Temperature Range: 0-100°C (32-212°F)
Pressure
Pressure
Minimum Conductivity: 100 μS/cm
Description and Specifications | 3 |