| Troubleshooting Guide |
| |
| No power LED or screen | Check power source, power switch (located on rear |
|
| does not come up. | of unit). |
|
| Pulse exhibits excessive | Check load impedance. |
|
| ringing. |
|
|
| Frequency limits. | Check duty cycle of EACH channel, both channels |
|
|
| must be within the specifications of the unit to |
|
|
| increase the frequency. Note that duty cycle of a |
|
|
| channel is the addition of the pulse width and the |
|
|
| delay. |
|
| Pulse Width limits. | Check the frequency and duty cycle. Note that duty |
|
|
| cycle of a channel is the addition of the pulse width |
|
|
| and the delay settings. |
|
| Pulse Delay limits. | Check the frequency and duty cycle. Note that duty |
|
|
| cycle of a channel is the addition of the pulse width |
|
|
| and the delay settings. |
|
| In external trigger pulse is | Check duty cycle of input trigger signal, check that |
|
| not stable or consistent | the signal frequency and trigger signal duty cycle are |
|
| with settings. | within specification. |
|
| In Burst Mode pulse is | Check to make sure timing constraints are not |
|
| not stable or consistent | exceeded; duty cycle of frequency is not exceeded |
|
| with settings. | with the addition of extra pulses. |
|
| Configuration will not | Make sure a valid configuration had been previously |
|
| load. | saved in configuration slot. |
|
| Sync output not a square | Check sync lines termination; note that sync does |
|
| wave or has excessive | not have same impedance as output lines. |
|
| ringing. |
|
|
Page 17