Barco Fusion 4MP manual ALC & Dicom Options, You can select the following DFs, Dicom offset

Page 28

Display information

You can select the following DFs:

Name

Description

 

 

DICOM

Select a DICOM display function for most

 

medical viewing applications.

 

The DICOM function results in more visi-

 

ble grayscales in the images.

 

 

Dyn Gamma 2.2

These are gamma functions that are

Dyn Gamma 1.8

shifted to take into account the non-zero

 

luminance of an LCD panel when driven

 

with a “black” signal.

 

They are especially useful in CT applica-

 

tions to improve the perception of low

 

Haunsfield values.

 

 

Native

If you select Native, the native panel

 

behavior will not be corrected.

 

 

Test

This is identical to the Native DF.

 

 

Gamma 2.2

Select one of these display functions in

Gamma 1.8

case the display is to replace a CRT dis-

 

play with a gamma of 1.8 or 2.2 respec-

 

tively.

 

 

User

This is a user-definable DF.

 

 

ALC & DICOM Options

DICOM offset

Introduction

The DICOM DF will be recalculated only when another DICOM Offset is selected. The DICOM DF will be recalculated based on the new offset setting.

To select a DICOM offset:

1.Select the DICOM display function. See To select a display function (DF):, on page 27.

28

Image 28
Contents Nio Fusion 4MP This page intentionally left blank Sources of information User interfaceOverview of product documentation Items in the navigation frameItems in the toolbar frame Graphic board control panel Languages supportedAccessing the Barco tab Using the Barco tabStart Settings Control Panel Display Barco Driver PropertiesDisplay Information Barco system reportBarco Adapter Configuration Advanced properties for Mxrt controllerMediCal QAWeb Display LayoutDisplay Resolution Bit TrueColor Mode Bit TrueColor Mode HDRColor Format Clone Primary display is cloned to the secondary displayMain Settings Custom SettingsOpenGL and Direct3D Wait for Vertical SyncDevice Details RotationWait for Vertical Sync Activating a configuration profile Modifying the configuration parametersOpenGL Profiles Adding a new application profileConfiguring the Windows desktop Graphic board control panel Precautions Optimize the lifetime of your displayUse a screen saver to avoid image retention Understand pixel technologyEnhance user comfort Maximize Quality AssuranceDisplay controls Overview of controlsFront controls are soft touch keys To move down or decrease values in the OSDHow to use the controls To put the display in standby modeTo switch from on state to stand-by Stand-by switchingTouch the Stand-by touch key Info messages Complete OSD On-screen display OSDLuminance and color menu Display Function menu Reading room definition submenu Settings menu Input Switch Mode Select Auto if you wish the display to Changing Display Functions Information menuConcepts InformationDisplay function selection To select a display function DFALC & Dicom Options You can select the following DFsDicom offset IntroductionYou can select the following offsets To select a reading room settingUse Enter to confirm the selection You can select the following reading room types To modify a reading room definitionSwitching Dpms on/off To switch Dpms on/offSwitching Power LED on/off To switch the power LED on/offTo disable the user controls Enabling/disabling User ControlsTo enable the user controls again Barco web sites Visit Barco at Visit Medical Imaging atNovember