EMC

This instrument has been designed to meet the requirements of the EMC Directive 89/336/EEC. Compliance was demonstrated by meeting the test limits of the following standards:

Emissions

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test limits used were:

a)Radiated: Class B

b)Conducted: Class B

c)Harmonics: EN61000-3-2 (2000) Class A; the instrument is Class A by product category.

Immunity

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use.

Test methods, limits and performance achieved were:

a)EN61000-4-2 (1995) Electrostatic Discharge : 4kV air, 4kV contact, Performance A.

b)EN61000-4-3 (1997) Electromagnetic Field, 3V/m, 80% AM at 1kHz, Performance B.

c)EN61000-4-11 (1994) Voltage Interrupt, 1 cycle, 100%, Performance B.

d)EN61000-4-4 (1995) Fast Transient, 1kV peak (AC line), 0.5kV peak (DC Outputs), Performance A.

e)EN61000-4-5 (1995) Surge, 0.5kV (line to line), 1kV (line to ground), Performance B.

f)EN61000-4-6 (1996) Conducted RF, 3V, 80% AM at 1kHz (AC line only; DC Output connections <3m not tested), Performance A.

According to EN61326 the definitions of performance criteria are:

Performance criterion A: ‘During test normal performance within the specification limits.’

Performance criterion B: ‘During test, temporary degradation, or loss of function or performance which is self-recovering’.

Performance criterion C: ‘During test, temporary degradation, or loss of function or performance which requires operator intervention or system reset occurs.’

Where Performance B is stated it is because DC Output regulation may deviate beyond Specification limits under the test conditions. However, the possible deviations are still small and unlikely to be a problem in practice.

Note that if operation in a high RF field is unavoidable it is good practice to connect the PSU to the target system using screened leads which have been passed (together) through an absorbing ferrite sleeve fitted close to the PSU terminals.

Cautions

To ensure continued compliance with the EMC directive observe the following precautions:

a)after opening the case for any reason ensure that all signal and ground connections are remade correctly and that case screws are correctly refitted and tightened.

b)In the event of part replacement becoming necessary, only use components of an identical type, see the Service Manual.

5

Image 6
Contents XPH Series 18V 10A DC Power Supply XPHPart Number Contact InformationTable of Contents IntroductionSpecification OutputSafety Emissions ImmunityMains Operating Voltage InstallationConnections Mains LeadOperation Series or Parallel Connection with Other Outputs MaintenanceCleaning VentilationAccess to Calibration Adjustments CalibrationVoltage Calibration Current CalibrationVoltage Regulation Sécurité AVERTISSEMENT! CET Instrument Doit Etre Relie a LA TerreConnexions Tension d’utilisation secteurCâble secteur Fonctionnement Connexion en série ou en parallèle avec d’autres sorties FusibleNettoyage Sicherheit WARNUNG! Dieses Gerät Muss Geerdet WerdenAnschlüsse NetzbetriebsspannungNetzkabel Betrieb VerbraucheranschlußIstwert-Fernerfassung Reihen- und Parallelschaltung mit anderen Ausgängen Wartung SicherungReinigung Sicurezza Attenzione Questo Strumento Deve Essere Collegato a TerraTensione d’esercizio InstallazioneCollegamenti Cavo d’alimentazioneFunzionamento Ventilazione ManutenzioneCollegamento in serie/parallelo con altre uscite FusibileSeguridad ADVERTENCIA! Este Instrumento Debe Conectarse a TierraTensión de la Red Eléctrica InstalaciónConexiones Cable de RedOperación Ventilación MantenimientoConexión en Serie o en Paralelo a Otras Salidas LimpiezaWarranty Information How do you get service?What does this warranty not cover? ProductReturn Material Authorization Policy Xantrex Technology Inc