Agilent Technologies DP105 Environmental and Physical, Operating Temperature, Relative Humidity

Page 40

3.11.2. Environmental and Physical

The modules have a Declaration of Conformity according to ISO/IEC Guide 22 and EN45014 and CE Marks of Compliance.

The American Standard MIL-PRF-28800F has been used as the source for many of the characteristics specified below. Those indicated with a ‘*’ are compatible with Class 3 as defined in that standard.

The front panels of the CompactPCI modules are in compliance with the IEEE 1101.10 standard.

Operating Temperature

0° to 40°C – for the DP1400 and DC271-FAMILY modules 0° to 50°C – all other modules

The above values are for the ambient temperature of the room (or equivalent) where the PC or CompactPCI crate is located. The temperature as measured on the board may well be significantly higher. On-board temperatures above 60°C should be avoided.

Relative Humidity*

5 to 95% (non-condensing)

Dimensions

All DP modules conform to the PCI standard. The DP214/DP235/DP240 and DP211/DP212/DP111 modules are 107

mm× 210 mm and require a long card slot. If used with the XP103 fan they can be adjusted to conform to the full

342 mm length. The DP1400 measures 107 mm × 170 mm without its fan. If used with the XP104 fan it can be adjusted to conform to the full long card length of 342 mm; when the XP105 fan is used the overall length is 235 mm. All other cards require a short card slot (107 mm × 175 mm). An XP100 fan will add 45 mm to the length of any card.

The DC110, DC140, and DC135 conform to the 3U CompactPCI standard (100 mm × 160 mm × 20 mm). All other DC modules conform to the 6U CompactPCI standard (233 mm × 160 mm × 20 mm).

Shock*

30 G, half-sine pulse

Vibration*

5-500 Hz, random

Safety

Complies with EN61010-1

EMC Immunity

Complies with EN61326-1: Industrial Environment

EMC Emissions

Complies with EN61326-1: Class A for radiated emissions

Required Airflow

> 2 m/s in situ

User Manual: Family of 8-bit Digitizers

Page 40 of 66

Image 40
Contents Agilent Acqiris Family BIT Digitizers XZ001USA Contents OUT of the BOX Running the Acqirislive Application Message to the User Using this ManualConventions Used in This Manual Disclaimer and Safety Packaging and HandlingWarranty System RequirementsMaintenance Transport & ShippingCleaning Disposal and RecyclingIC414 Installation 1. IC414 Hardware installation hints 2. IC414 Windows software installationInstallation Installing the Software under WindowsMultiple Versions Page Page Page Page Page Page Page Page Page Installing the Software for Linux Kernel Mode Driver Compilation ƒ cd linuxdriverpciSpecial cases Installing the HardwareEnvironment variables for the Firmware After Restarting Windows Windows XPPage LabVIEW RT Installing the IVI-COM/C Driver Distribution for Windows 2000/XP and LinuxDP110 Block Diagram OverviewChannel Input Coupling ImpedanceInput Protection Mezzanine Front-endOffset Particularities Input Voltage and Offset Vertical ResolutionDC Accuracy and Linearity Using Probes Data AcquisitionAcquisition Memory Sampling RateSingle and Sequence Acquisition Modes Timing Timebase RangeCombining channels TriggerTrigger Coupling Trigger LevelEdge Trigger Slope Window TriggerTrigger Status Pre- and Post-Trigger DelayExternal Clock and Reference ASBus Internal CalibrationSpecial Front Panel Input and OutputDP1400 & DC271-FAMILY Frequency Counter mode External Trigger Output DP1400 & DC271-FAMILYPCI DMA Operating Temperature DimensionsSafety Environmental and PhysicalRunning the AcqirisLive Application Getting Started with AcqirisLiveControl Panel Mode Control Panel and FunctionsDisplaying Multiple Traces Keyboard shortcuts Numeric EditorInput Voltage Scale, Offset and Coupling Acquisition Mode Using digitizers of different models simultaneously Timebase and MemoryDisplay Features, Zoom and Persistence Segmented MemoryTop Line Menu of AcqirisLive SetupStore, Autostore and Conversion of Waveforms Page Calibrate Combine channelsExternal 10 MHz Reference and External Clock Additional Waveform InformationDisplay Features Command Line SwitchesAvailable Switches Operation ModeTransient Recorder mode Buffer sizeBanks for Smar CalibrationSimulation Mode Temperature UpdateAcquisition State Multi-instrumentsHow to run GeoMapper When should the GeoMapper Application be usedWho needs a Geographical Map of Modules Page Page Page Appendix B ASBus2 for PCI Digitizers Page Page Page Page Page Appendix H XP104 Fan Unit for DP1400 Page Appendix I XP105 Fan Unit for DP1400