Agilent Technologies DP106, DP240 Who needs a Geographical Map of Modules, How to run GeoMapper

Page 54

5.Running the GeoMapper Application

5.1.Who needs a Geographical Map of Modules

The GeoMapper application gives the user the possibility of telling the Acqiris driver crucial details of the physical position of modules when this information cannot be deduced directly. It allows the driver to correctly adjust trigger and clock delays between the different modules of an ASBus multi-Instrument and arrange the modules in an intuitive order.

The GeoMapper application has been implemented as a Wizard with the goal of learning the physical position of each module of a multi-bus system. The driver already knows the logical Bus and logical Slot numbers of each Acqiris unit present. However, because of varying conventions for the compactPCI slot arrangement on a bus, this is not sufficient to determine the physical position. Furthermore, with the aid of the map, the software can ensure that the modules of an Auto-defined multi-Instrument will be ordered to correspond to increasing physical slot number.

NOTE: The unusual slot numbering convention of the CC121 is documented in the CC121 CompactPCI Crate User Manual. The CC10X crates have a simpler slot numbering convention that the Acqiris driver already takes into account. CompactPCI Crates from other manufacturers represent an unknown element.

NOTE: The GeoMapper is of use for some ASBus systems in CC121 crates and, perhaps, crates from other manufacturers. CC121 crate configurations with both an Acqiris acquisition module in one of the last 7 slots and with the PC running under Windows 2000 or XP will not need the GeoMapper. Users who do not need GeoMapper do not need to read this chapter any further.

5.2.When should the GeoMapper Application be used

The application needs to be used whenever the bus or slot configuration of the system changes in a fundamental way. The most obvious trouble symptom is that an expected multi-Instrument was not found. Alternatively the driver can issue error messages that the AcqirisLive application and other programs will see. Among them are

Modules are not contiguous in ASBus combination ASBus Candidate Modules not on same bus 'AcqGeo.map' file is invalid or corrupted

ACQIRIS_ERROR_MODULES_NOT_CONTIGUOUS ACQIRIS_ERROR_MODULES_NOT_ON_SAME_BUS ACQIRIS_ERROR_INVALID_GEOMAP_FILE

In all of these cases the GeoMapper application should be run to ensure that the map file is up to date.

If desired the GeoMapper application will generate a default configuration in which it allows the driver to make all of the choices. This is done by specifying "None (Disable GeoMap)" as the desired configuration in "Step1: Crate count" of the application.

5.3.How to run GeoMapper

After software installation you will find GeoMapper in the AcqirisApp directory of the installation directory. After being started the program will learn about the Acqiris devices present and then show a screen like the one below:

The choice of the number of crates ranges from None up to the number of bus segments that the driver has found containing Acqiris modules. If you choose None and save the resulting AqGeo.map file the system will behave in its natural way without any additional user provided information. If you want to give the driver information on the physical position of modules you should select the appropriate number of crates. If the number of crates does not

User Manual: Family of 8-bit Digitizers

Page 54 of 66

Image 54
Contents Agilent Acqiris Family BIT Digitizers XZ001USA Contents OUT of the BOX Running the Acqirislive Application Using this Manual Message to the UserConventions Used in This Manual Disclaimer and Safety Packaging and HandlingWarranty System RequirementsMaintenance Transport & ShippingCleaning Disposal and RecyclingIC414 Installation 1. IC414 Hardware installation hints 2. IC414 Windows software installationInstalling the Software under Windows InstallationMultiple Versions Page Page Page Page Page Page Page Page Page Installing the Software for Linux Kernel Mode Driver Compilation ƒ cd linuxdriverpciInstalling the Hardware Special casesEnvironment variables for the Firmware After Restarting Windows Windows XPPage LabVIEW RT Installing the IVI-COM/C Driver Distribution for Windows 2000/XP and LinuxDP110 Block Diagram OverviewInput Protection Channel Input CouplingImpedance Mezzanine Front-endOffset Particularities Vertical Resolution Input Voltage and OffsetDC Accuracy and Linearity Using Probes Data AcquisitionSampling Rate Acquisition MemorySingle and Sequence Acquisition Modes Combining channels TimingTimebase Range TriggerEdge Trigger Slope Trigger CouplingTrigger Level Window TriggerPre- and Post-Trigger Delay Trigger StatusExternal Clock and Reference Internal Calibration ASBusSpecial Front Panel Input and OutputDP1400 & DC271-FAMILY Frequency Counter mode External Trigger Output DP1400 & DC271-FAMILYPCI DMA Safety Operating TemperatureDimensions Environmental and PhysicalRunning the AcqirisLive Application Getting Started with AcqirisLiveControl Panel Mode Control Panel and FunctionsDisplaying Multiple Traces Numeric Editor Keyboard shortcutsInput Voltage Scale, Offset and Coupling Acquisition Mode Using digitizers of different models simultaneously Timebase and MemoryDisplay Features, Zoom and Persistence Segmented MemorySetup Top Line Menu of AcqirisLiveStore, Autostore and Conversion of Waveforms Page External 10 MHz Reference and External Clock CalibrateCombine channels Additional Waveform InformationAvailable Switches Display FeaturesCommand Line Switches Operation ModeBanks for Smar Transient Recorder modeBuffer size CalibrationAcquisition State Simulation ModeTemperature Update Multi-instrumentsWhen should the GeoMapper Application be used How to run GeoMapperWho needs a Geographical Map of Modules Page Page Page Appendix B ASBus2 for PCI Digitizers Page Page Page Page Page Appendix H XP104 Fan Unit for DP1400 Page Appendix I XP105 Fan Unit for DP1400