Agilent Technologies DC211, DP240, DP212, DP214 Control Panel and Functions, Control Panel Mode

Page 42

Other controls in the control panel allow you to select the optimal acquisition settings for your particular signal. Each of the other control panel functions will be discussed in detail in the following Control Panel Functions section. Here is a sample result.

4.2.Control Panel and Functions

DSO TR

4.2.1. Control Panel Mode

 

In AcqirisLive there are two control panel modes available, Oscilloscope Mode and Transient

 

Recorder Mode. The choice of mode is entirely determined by the preference of the user.

Both modes provide the same set of functional capabilities but display the settings slightly differently. For example in Transient Recorder Mode, the user selects the Full Scale Range of the amplifier whereas in Oscilloscope Mode the user selects the input voltage scale set in Volts/Div. The desired mode is selected using one of the two buttons shown at the top of the control panel. The distinction between the control panels for the two operating modes can be seen in the images below.

In both modes it is possible to set the memory depth and the sample speed independently within the limits of the maximum available memory. However the two modes differ in the following way.

In Oscilloscope Mode, the user selects the time window (time per div) and the sampling rate will be adjusted to the fastest possible within the limits of the available memory. The maximum available memory is programmable.

In Transient Recorder Mode, the user selects the capture time window, the sampling rate, or the number of samples. When selecting the capture time window, the number of samples is adapted, keeping the current sampling rate and number of segments, up to the maximum available memory on the card. When selecting the sampling rate, the number of samples is adapted, keeping the current capture time window and number of segments, up to the maximum available memory on the card. When selecting the number of samples, the capture time window is adapted, keeping the current sampling rate and number of segments.

User Manual: Family of 8-bit Digitizers

Page 42 of 66

Image 42
Contents Agilent Acqiris Family BIT Digitizers XZ001USA Contents OUT of the BOX Running the Acqirislive Application Using this Manual Message to the UserConventions Used in This Manual Disclaimer and Safety Packaging and HandlingWarranty System RequirementsMaintenance Transport & ShippingCleaning Disposal and RecyclingIC414 Installation 1. IC414 Hardware installation hints 2. IC414 Windows software installationInstalling the Software under Windows InstallationMultiple Versions Page Page Page Page Page Page Page Page Page Installing the Software for Linux Kernel Mode Driver Compilation ƒ cd linuxdriverpciInstalling the Hardware Special casesEnvironment variables for the Firmware After Restarting Windows Windows XPPage LabVIEW RT Installing the IVI-COM/C Driver Distribution for Windows 2000/XP and LinuxDP110 Block Diagram OverviewInput Protection Channel Input CouplingImpedance Mezzanine Front-endOffset Particularities Vertical Resolution Input Voltage and OffsetDC Accuracy and Linearity Using Probes Data AcquisitionSampling Rate Acquisition MemorySingle and Sequence Acquisition Modes Combining channels TimingTimebase Range TriggerEdge Trigger Slope Trigger CouplingTrigger Level Window TriggerPre- and Post-Trigger Delay Trigger StatusExternal Clock and Reference Internal Calibration ASBusSpecial Front Panel Input and OutputDP1400 & DC271-FAMILY Frequency Counter mode External Trigger Output DP1400 & DC271-FAMILYPCI DMA Safety Operating TemperatureDimensions Environmental and PhysicalRunning the AcqirisLive Application Getting Started with AcqirisLiveControl Panel Mode Control Panel and FunctionsDisplaying Multiple Traces Numeric Editor Keyboard shortcutsInput Voltage Scale, Offset and Coupling Acquisition Mode Using digitizers of different models simultaneously Timebase and MemoryDisplay Features, Zoom and Persistence Segmented MemorySetup Top Line Menu of AcqirisLiveStore, Autostore and Conversion of Waveforms Page External 10 MHz Reference and External Clock CalibrateCombine channels Additional Waveform InformationAvailable Switches Display FeaturesCommand Line Switches Operation ModeBanks for Smar Transient Recorder modeBuffer size CalibrationAcquisition State Simulation ModeTemperature Update Multi-instrumentsWhen should the GeoMapper Application be used How to run GeoMapperWho needs a Geographical Map of Modules Page Page Page Appendix B ASBus2 for PCI Digitizers Page Page Page Page Page Appendix H XP104 Fan Unit for DP1400 Page Appendix I XP105 Fan Unit for DP1400