Sigma BM-286A service manual Operating Instructions

Page 22

3. OPERATING INSTRUCTIONS page 19

Alarm Type:

Alarm type selects the method of generating a radiation alarm.

The available options are:

1.NSIG: ((N*1background) + background)

2.DPM: Disintegrations per minute above background

3.SPACT: Specific Activity in pCi/gram

Alarm Message:

This parameter entry is the message that will be displayed if the alarm level is exceeded. This may be up to 22 alpha-numeric characters. This message will be included on the data sample information and in the reports.

Clean Message:

This parameter entry is the message that will be displayed if the alarm level is not exceeded. This may be up to 22 alpha-numeric characters. This message will be included on the data sample information and in the reports.

Maximum Weight:

Enter a maximum allowable weight, in pounds. This is used to determine if the item to be scanned is within the waste stream weight limits.

Model BM-286A

Image 22
Contents Waste Monitor Model BM-286A Maintenance Table of ContentsInspection and Installation Theory of OperationConfiguration Tracking Sheet Parts TroubleshootingAppendix Warranty Glossary Formulas Recommended Spare Parts Spare Parts Ordering InformationIntroduction Scope and Purpose of ManualSpecifications Introduction Introduction Inspection Installation Operating Instructions General OverviewPOWER-UP and SELF-TESTS Access Menu Level One AccessPCA-II Normal Operation Operating Instructions Programming the System SET-UP Menu AbcdefghOperating Instructions Operating Instructions Waste Stream Parameters Operating Instructions Operating Instructions Operating Instructions This value is used only when the Alarm Type is set to DPM Utilities Menu Operating Instructions Reports Initial Installation Checklist Operating Instructions BM-286A Waste Stream SET-UP Worksheet Unit Operation Theory of Operation Theory of Operation Theory of Operation Theory of Operation Operator Maintenance Maintenance Maintenance LLD Settings for Plastic DetectorsULD Weight Calibration Maintenance Diagnostic Aids Fault Isolation Computers Longs Peak Court LONGMONT, Colorado Appendix B. Glossary Appendix B. Glossary Activity from Counts Reliable Detectable Activity RDA Formula Variance Configuration Tracking Sheet System ModificationsLongs Peak Court LONGMONT,COLORADO Drawings