5. MAINTENANCE page 35

Connect the DVM between TP16(-) and TP17(+).

Connect the oscilloscope between TP16(ground) and TP15(input).

Place 137Cs source as close to the center of the detector under test.

Refer to Drawing 20 (Appendix F.) for an example of a typical pulse profile.

Adjust the oscilloscope controls to provide a similar display.

Adjust R42 for 1.0 volts peak pulse amplitude.

Verify that TP17 is <14 volts.

Set oscilloscope: 2 volts VOLTS\DIV

Move oscilloscope probe to TP13. Adjust R5 for 4 volts peak pulse amplitude.

NOTE:

Pulses of all amplitudes will be observed on the scope. Refer to the typical pulse profile

on Drawing 20 (Appendix F.). However there should be a noticeable increase in the display intensity for those pulses having amplitudes of 4.5 volts or less relative to those of greater amplitudes. If this is not readily apparent, try adjusting the scope intensity control.

Model BM-286A

Image 38
Contents Waste Monitor Model BM-286A Maintenance Table of ContentsInspection and Installation Theory of OperationConfiguration Tracking Sheet Parts TroubleshootingAppendix Warranty Glossary Formulas Recommended Spare Parts Spare Parts Ordering InformationIntroduction Scope and Purpose of ManualSpecifications Introduction Introduction Inspection Installation Operating Instructions General OverviewPOWER-UP and SELF-TESTS Access Menu Level One AccessPCA-II Normal Operation Operating Instructions Programming the System SET-UP Menu AbcdefghOperating Instructions Operating Instructions Waste Stream Parameters Operating Instructions Operating Instructions Operating Instructions This value is used only when the Alarm Type is set to DPM Utilities Menu Operating Instructions Reports Initial Installation Checklist Operating Instructions BM-286A Waste Stream SET-UP Worksheet Unit Operation Theory of Operation Theory of Operation Theory of Operation Theory of Operation Operator Maintenance Maintenance Maintenance ULD Settings for Plastic DetectorsLLD Weight Calibration Maintenance Diagnostic Aids Fault Isolation Computers Longs Peak Court LONGMONT, Colorado Appendix B. Glossary Appendix B. Glossary Activity from Counts Reliable Detectable Activity RDA Formula Variance Configuration Tracking Sheet System ModificationsLongs Peak Court LONGMONT,COLORADO Drawings