Sigma BM-286A service manual Computers

Page 44

6. TROUBLESHOOTING page 41

6.3. COMPUTERS

To check the DOS computer a VGA monitor can be connected to see if it is running and the output to the screen will look something like this:

(1,235,234,334,123,234,132)

The information that is being sent is door position and six detector counts.

The RS-232 data to the NT machine can be examined by using Hyperterminal. The settings should be:

COM1,9600,N,8,1

Use standard troubleshooting procedures on the NT machine.

6.4. SIGNAL INTERFACE ASSEMBLY AND DETECTORS

Channel boards (RCS-110) may be substituted with system power on, but the power must be turned off before changing the discriminator cards (RCS-120). This will ensure that the discriminator settings are updated.

NOTE:

If channel cards are replaced permanently, the channels affected must be recalibrated.

Model BM-286A

Image 44
Contents Waste Monitor Model BM-286A Table of Contents Inspection and InstallationMaintenance Theory of OperationTroubleshooting Appendix Warranty Glossary FormulasConfiguration Tracking Sheet Parts Recommended Spare Parts Spare Parts Ordering InformationIntroduction Scope and Purpose of ManualSpecifications Introduction Introduction Inspection Installation Operating Instructions General OverviewPOWER-UP and SELF-TESTS Access Menu Level One AccessPCA-II Normal Operation Operating Instructions Programming the System SET-UP Menu AbcdefghOperating Instructions Operating Instructions Waste Stream Parameters Operating Instructions Operating Instructions Operating Instructions This value is used only when the Alarm Type is set to DPM Utilities Menu Operating Instructions Reports Initial Installation Checklist Operating Instructions BM-286A Waste Stream SET-UP Worksheet Unit Operation Theory of Operation Theory of Operation Theory of Operation Theory of Operation Operator Maintenance Maintenance Maintenance ULD Settings for Plastic DetectorsLLD Weight Calibration Maintenance Diagnostic Aids Fault Isolation Computers Longs Peak Court LONGMONT, Colorado Appendix B. Glossary Appendix B. Glossary Activity from Counts Reliable Detectable Activity RDA Formula Variance Configuration Tracking Sheet System ModificationsLongs Peak Court LONGMONT,COLORADO Drawings